Using the iN∫X
5.12 Direct Power: Both Legs
In this mode, both sets of legs move when or is pressed.
Be aware:
●
Auto-equalize is off. Operators must maintain control of
the iN∫X. Do not put the patient at a severe angle.
●
The weight of the patient and equipment may cause the
legs to move at different speeds.
●
Display (Figure 48):
○
White background
○
Blue legs and double-headed arrows indicates both
legs will move.
○
Red wheels indicate that the iN∫X must not be rolled.
○
If the 15-second countdown timer expires, the iN∫X
returns to its standard operating mode.
5.13 Direct Power: Loading-End Legs
In this mode, only the loading-end legs move when or is
pressed. Be aware:
●
Auto-equalize is off. Operators must maintain control of
the iN∫X. Do not put the patient at a severe angle.
●
Display (Figure 49):
○
White background
○
Blue loading-end legs and double-headed arrow
indicates only the loading-end legs will move.
○
Red wheels indicate that the iN∫X must not be rolled.
○
If the 15-second countdown timer expires, the iN∫X
returns to its standard operating mode.
5.14 Direct Power: Control-End Legs
In this mode, only the control-end legs function when or
is pressed. Be aware:
●
Auto-equalize is off. Operators must maintain control of
the iN∫X. Do not put the patient at a severe angle.
●
Display (Figure 50):
○
White background
○
Blue control-end legs and double-headed arrow
indicates only the control-end legs will move.
○
Red wheels indicate that the iN∫X must not be rolled.
○
If the 15-second countdown timer expires, the iN∫X
returns to its standard operating mode.