11
5.7 Storing the Splints
Store clean, dry splints in the carrying case pockets. Slide the vacuum
pump into the pocket in the bottom of the carrying case.
Using the Splints
5.6 Removing the Splint
From the Patient
The vacuum pump is not used when removing the splint from the
patient.
1. One operator manually stabilizes the injured area of the body
throughout the splint removal procedure.
2. The second operator unfastens the adjustment straps.
3. Open the air valve by turning the valve knob counter-clockwise
to allow air into the splint. As air flows in, the splint will become
pliable (Figure 9). Notee:
You can completely unthread the valve
knob from the valve if desired (Figure 10).
4. When the splint is pliable, gently remove the splint while
stabilizing the injured area of the body.
9
10
Removable Valve Knob