![Ferno 35X-NM PROFlexx Скачать руководство пользователя страница 2](http://html.mh-extra.com/html/ferno/35x-nm-proflexx/35x-nm-proflexx_manual_554372002.webp)
© Ferno-Washington, Inc. / 234-3494-01 / July 2018
35X-NM Non-Magnetic Cot
2
MR Conditional Safety Label
Safety and instruction labels place important information from the
users’ manual on the cot. Read and follow label instructions. Replace
worn or damaged labels immediately. New labels are available from
Ferno. See “Ferno Customer Relations” on page 8.
Magnetic-Resonance Rating Label (Qty. 1)
See below for Listed Conditions of Use
MR Conditional Symbol
The symbol at right is used on the cot and in this users’ manual. Ferno
uses symbols recognized by the International Standards Organization
(ISO), American National Standards Institute (ANSI), the American
Society for Testing and materials (ASTM), and the emergency medical
services industry.
Model 35X-NM Cot is
Magnetic-Resonance Conditional
MR Conditional Testing and Compliance
TESTING
The Model 35X-NM cot has been tested and is listed as “MR Conditional”.
A product that is Magnetic-Resonance Conditional (called MR
Conditional in this manual and on product labeling) has been shown to
pose no hazards in a specified MR environment under listed conditions
of use.
LISTED CONDITIONS OF USE
The Model 35X-NM cot has undergone testing and has been determined
to be MR Conditional. In general, the cot is intended for use under the
conditions listed here:
●
Static magnetic field of 3-Tesla or less
●
Spatial gradient magnetic field of 720-Gauss/cm or less
●
Inside the MR environment (e.g., in the MR system room, close to
the scanner)
The cot is not designed or designated for use directly inside the MR
system (e.g., inside the bore of the scanner).
COMPLIANCE
Non-clinical magnetic field interaction testing has been conducted by
an independent, third-party testing service.
MR Conditional Label Affixed to Cot
NOTICE
The cot is not designed or designated for use directly inside
the MR system (e.g., inside the bore of the scanner).
ASTM Contact Information
ASTM International
100 Barr Harbor Drive, P.O. Box C700
West Conshohocken, PA 19428-2959, USA
Internet: