© Ferno-Washington, Inc. / 234-3494-01 / July 2018

35X-NM Non-Magnetic Cot

2

MR Conditional Safety Label

Safety and instruction labels place important information from the 

users’ manual on the cot. Read and follow label instructions. Replace 

worn or damaged labels immediately. New labels are available from 

Ferno. See “Ferno Customer Relations” on page 8.

Magnetic-Resonance Rating Label (Qty. 1)

See below for Listed Conditions of Use

MR Conditional Symbol

The symbol at right is used on the cot and in this users’ manual. Ferno 

uses symbols recognized by the International Standards Organization 

(ISO), American National Standards Institute (ANSI), the American 

Society for Testing and materials (ASTM), and the emergency medical 

services industry.

Model 35X-NM Cot is  

Magnetic-Resonance Conditional

MR Conditional Testing and Compliance

TESTING

The Model 35X-NM cot has been tested and is listed as “MR Conditional”.
A product that is Magnetic-Resonance Conditional (called MR 

Conditional in this manual and on product labeling) has been shown to 

pose no hazards in a specified MR environment under listed conditions 

of use.

LISTED CONDITIONS OF USE

The Model 35X-NM cot has undergone testing and has been determined 

to be MR Conditional. In general, the cot is intended for use under the 

conditions listed here:

● 

Static magnetic field of 3-Tesla or less

● 

Spatial gradient magnetic field of 720-Gauss/cm or less

● 

Inside the MR environment (e.g., in the MR system room, close to 

the scanner)

The cot is not designed or designated for use directly inside the MR 

system (e.g., inside the bore of the scanner).

COMPLIANCE

Non-clinical magnetic field interaction testing has been conducted by 

an independent, third-party testing service.

MR Conditional Label Affixed to Cot

 

NOTICE

The cot is not designed or designated for use directly inside 

the MR system (e.g., inside the bore of the scanner).

ASTM Contact Information

ASTM International
100 Barr Harbor Drive, P.O. Box C700
West Conshohocken, PA  19428-2959, USA
Internet: 

https://www.astm.org/

Содержание 35X-NM PROFlexx

Страница 1: ...equirements of the Star of Life certification via Federal Ambulance Specification KKK A 1822 For information contact Federal Supply Services Specifications Section Suite 8100 470 E L Enfant Plaza SW Washington DC 20407 Purpose of This Manual This manual Ferno document 234 3494 is a supplement to the 35X PROFlexx Series Users Manual The specific features of the 35X NM cot are described in this docu...

Страница 2: ... product that is Magnetic Resonance Conditional called MR Conditional in this manual and on product labeling has been shown to pose no hazards in a specified MR environment under listed conditions of use LISTED CONDITIONS OF USE The Model 35X NM cot has undergone testing and has been determined to be MR Conditional In general the cot is intended for use under the conditions listed here Static magn...

Страница 3: ... 3 Slide the looped end of a strap around the selected attachment point and thread the buckle or loose end of the strap through the loop Figure 2 Pull the loop tight around the attachment point 4 Repeat Steps 1 3 with the remaining five straps 5 Fasten the restraints USING CAM BUCKLE PLASTIC RESTRAINTS To fasten a cam buckle restraint Open the cam lock and thread the strap through the opening Figu...

Страница 4: ... locked into the new position After Use Buckle and or adjust the torso restraint Sidearms Purpose Sidearms provide patient security and comfort Keep the sidearms raised except when transferring a patient to or from the cot The sidearms feature an intermediate stop that allows a patient s arm to be placed on the sidearm for treatment starting of intravenous IV fluids etc Use To lower pull the relea...

Страница 5: ...Cot Before transferring the patient to the cot confirm the style of restraint on the cot is correct for the cot s use See Patient Restraints on page 3 Use only the plastic cam buckle restraints provided when the cot will be used in an MR environment Use only primary restraints such as metal buckle restraints when the cot will be used in an ambulance Figure 7 Wheel Lock NOTICE Different styles of p...

Страница 6: ...the cam buckle restraint buckles and webbing following the disinfectant manufacturer s directions for application method and contact time Cleaning Place the cam buckle restraints in a mesh bag not supplied and machine wash in warm water with a mild detergent Hang the restraints to dry Do not place the restraints in a dryer METAL BUCKLE OR OTHER RESTRAINTS Disinfect and clean only as directed in th...

Страница 7: ... tubing Do not lubricate Wheels transport and loading Do not lubricate Center leg crosstube Do not lubricate Backrest hinge Do not lubricate Telescoping side handles Do not lubricate Telescoping legs Do not lubricate Shock frame hinge Do not lubricate Control handle mechanisms Do not lubricate Applying Lubricant Use a small amount of lubricant one or two drops Lubricate identical points on each si...

Страница 8: ...itions of the limited warranty and the limitations of liability and disclaimers are also available upon request by calling Ferno at 1 800 733 3766 or 1 937 382 1451 Repairs U S A and Canada EMSAR istheonlyorganizationauthorizedbyFerno to manage service and repair Ferno products Telephone Toll free 1 800 73 EMSAR Telephone Worldwide 1 937 383 1052 Fax Worldwide 1 937 383 1051 Internet www EMSAR com...

Отзывы: