Maintenance
73
MY INNE READER IS SHOWING A RED UI LIGHT
RED STATIC LIGHT
If the inne READER shows a solid red UI light,
it means the device has encountered an error. The
error could either be a too low battery level or an
error when analysing your saliva sample. In order to
identify the error type please follow the steps below
1. Check the inne READER’s battery level:
Press the button on the inne READER shortly
and check the battery light of the inne
READER.
→ RED BATTERY LIGHT or NO BATTERY LIGHT:
Battery level is too low to take a
measurement.
STEPS: Connect the inne READER to a
charger and wait till the battery light
changes from red to yellow. Then take a
new inne STRIP and repeat the progesterone
measurement.
→ YELLOW / GREEN BATTERY LIGHT:
Battery level is sufficient to do a
reading. Problem with the collected sample
or the inne STRIP.
STEPS: Repeat the test with a new inne
STRIP to get the daily progesterone
reading. Follow the steps outlined in p.48
RED BLINKING LIGHT
If the inne READER shows you a blinking red
light and goes off afterwards, it means that it has
encountered a system error. Please contact customer
service by sending an email to [email protected]