Fema C Series Скачать руководство пользователя страница 9

FEMA ELECTRÓNICA . Series C . C40-D

9

1.13 Measuring with Pt and Ni probes

 

Accepted Pt and Ni probes

The instrument accepts connection of Pt100, Pt500 and 

Pt1000 temperature probes, and also Ni100, Ni200 and 

Ni1000  temperature  probes.  Temperature  ranges  for 

each type of probe are indicated on 

‘Table 11’

 below.

 

Pt100 with 2 and 3 wires

The instrument accepts connection for 2 and 3 wire Pt100 probes. 

For 3 wire Pt100 probes, configure internal jumper ‘T’ at position 5-6 

(see section 1.9). 

For 2 wire Pt100 probes, configure internal jumper 

‘T’ at position 4-5. Compensation of the wire resistance for 2 wire 

probes can be manually configured with the parameter ‘

Offset read-

ing

’ (‘

oFFS

’) 

(see section 1.22)

 which allows to configure a fixed num

-

ber of counts to be added to the reading.

 

Resolution and units

The temperature resolution using Pt and Ni temperature probes is 

configurable to 1º or 0.1º. Reading can be configured in ºC (degrees 

Celsius) or ºF (degrees Fahrenheit).

 

Sensor break detection

In case of sensor break, the instrument will show ‘

h.ovr

’ or ‘

h.udr

’ 

(see section 1.28) 

depending on the broken cable.

 

Alpha temperature coefficient

The instrument accepts Pt probes (Pt100, Pt500, Pt1000) with ‘

Al-

pha

’ parameter of ‘

0.0385

’ and ‘

0.0390

’ 

(see section 1.29.7)

This pa

-

rameter is associated to the specific model of probe installed.

Sensor

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Range in ºC

 

(in ºF)

Total error

Connection

 

(

terminals)

Current at 

sensor

Pt100 3 wires

F H J

5-6

-200 / 700 ºC 

(-328 / 1292 ºF)

<1º

2 (Pt+) 

4 (Pt -) 

5 (sense) 

< 900 uA

Pt100 2 wires

F H

4-5

-200 / 700 ºC 

(-328 / 1292 ºF)

2 (Pt+, Ni+) 

4 (Pt -, Ni-)

Pt500

F

-150 / 630 ºC 

(-238 / 1166 ºF)

< 90 uA

Pt1000

F

-190 / 630 ºC 

(-310 / 1166 ºF)

< 90 uA

Ni100

F H

-60 / 180 ºC 

(-76 / 356 ºF)

< 900 uA

Ni200

F H

-60 / 120 ºC 

(-76 / 248 ºF)

< 900 uA

Ni1000

F

-60 / 180 ºC 

(-76 / 356 ºF)

< 90 uA

Table 11 - Ranges of temperature for Pt and Ni probes

 

Terminal 5 ‘multifunction’ - ‘Pt100 3 wire’ or ‘External control’

To configure for 3 wire Pt100 probes, set internal jumper ‘T’ at posi

-

tion 5-6 

(see section 1.9). 

Terminal 5 will be assigned to the connec

-

tion of the third wire of the Pt100 probe, for automatic wire resis

-

tance compensation purposes, up to 30 R.
To configure for 2 wire Pt100 probes, set internal jumper ‘T’ at posi

-

tion 4-5 

(see section 1.9). 

Terminal 5 will be configured as ‘EK’ exter

-

nal control function. See section 

1.9 

for a list of available functions.

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal 

connections are indicated at section 

1.5.

 Location for internal jump

-

ers is indicated at section 

1.9.

See below a list of typical connections :

Pt+, Ni+

Pt-, Ni-

1 2 3 4 5

 

other probes 2 wires

 

 

Pt100 3 wires

 

Pt-

Sense

Pt+

1 2 3 4 5

Содержание C Series

Страница 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Страница 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Страница 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Страница 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Страница 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Страница 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Страница 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Страница 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Страница 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Страница 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Страница 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Страница 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Страница 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Страница 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Страница 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Страница 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Страница 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Страница 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Страница 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Страница 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Страница 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Страница 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Страница 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Страница 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Страница 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Страница 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Страница 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Страница 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Страница 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Страница 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Страница 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Страница 32: ......

Отзывы: