F-940 Instruction Manual rev. 3-8-2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
38
Phone: (360) 833-8835
[email protected]
www.felixinstruments.com
3.
Unplug the red and black cable connector by gently
pulling the fastener out the end of the O₂ sensor.
4.
Unscrew the O
2
sensor (counter-clockwise).
5.
Screw in the new sensor (clockwise) until you feel a
resistance
—
not too tight!
6.
Plug in the cable connector by pressing the white
fastener into the end of the new O2 sensor.
7.
Screw the bottom plate into position, and fasten the
battery cap.
8.
Place the bottom rubber mat
—
and congratulations
you’ve done it!
9.
Photos of the process are below.