Date & Time
MM/DD/YYYY
hh : mm : ss
04/21/2016
17 : 01 : 43
Setup > RH Conversion
The Setup RH Conversion menu is used to
correct for the temperature sensor being inside
F-940 housing. This causes the temperature
sensor to reflect the temperature of the F-940
and not the ambient temperature, which is
usually several degrees above ambient. The
Relative Humidity is calculated based on the
temperature value.
RH Conversion
Enter Sample T(C)
20.0
Use Sample T
Yes
(for measurement)
Device RH (%)
43.6
Sample RH (%)
21.3
Enter the correct ambient temperature in
degrees Celsius, measured from an external
temperature sensor. To use the entered
temperature for RH calculations
, “use sample
T
” should be set to “Yes”.
The current (internal RH) and corrected
(sample RH) are displayed at the bottom of the
screen.