DIGITAL  WATER BATH  OPERATION

a) Auto-revision and turn on of the system

The Felisa Digital Water Bath has an error auto-check  that let you know from the most common failures 
of the system, therefore   the user can be sure that the Water Bath is working in optimal conditions. 
Every time you turn on the equipment, the system auto test most of its components. This test is a 
revision of the crucial features of the system.
The display present for two seconds the number “8888” at the same time, the audible alarm is activated. 
After this, the display turns off and the system starts the internal test. If all the features are all right then 
the Water Bath  will present in the display the water temperature. If any of the features fail, the display  
will present the message “ErrX”. Where the X is replaced by a number from 0 to 7 depending of the error 
detected. The audible alarm will turn on. Press the increase button to restart, if the error persist   you 
should contact or send the equipment  to our Plant  or any authorized Distributor.

Failure Code:
Sensor Failure

                      Err (0 y 4)

Heating Element Failure

           Err (2)

b) Setting the working temperature

The Water Bath has a simple menu that allows you to configure the working  mode of the system. This 
menu is accessible when the key switch (31)   is in the open position, otherwise you won`t be able to 
change the configuration that is established at that moment.
Press the menu button, the display will present the message “SP” (Set Point)  for two seconds and then 
the configure temperature. With the increase or decrease buttons you can modify this parameter to the 
desired operation temperature, then press the menu button to continue with the configuration and to 
record the selected value.

c) Setting  the overshoot alarm

At this point the display will present the message “SPHA” and then will present the temperature of  the 
overshoot alarm. This temperature will always be  1 °C higher than the set  point temperature.
To modify the overshoot alarm use the increase or decrease buttons to set the desired temperature, 
and then press the menu button to record  and   continue with the configuration.
 

d) Setting the Timer

The Water Bath has a 99 hours 59 minutes capacity timer   which can be activated according to your 
needs.
When you finish the configuration of the overshoot alarm, the display will present the message “ tr “ an 
then the present configuration of the timer ( “On” or  “Off” ).
If you want to work with the timer press the increase button to select the “On” option, if you choose to 
work without the timer press the decrease button  to select the “Off” option. When you have done your 
selection press the menu button to record and continue with the configuration. If you active the timer, 
the menu will show the setting of the hours, if not you will go directly to the configuration of the audible 
alarm. 

Setting the hours in the timer: the display will present the message “ hh ‘. To adjust the hours use the 
increase or decrease buttons, you can go from 0 to 99 hours. When you finish the selection of the hours 
press the menu button to record the new configuration in the system. Then the display will present the 

Содержание FE-371

Страница 1: ...ontrol an logico o con control digital todos los modelos tienen el elemento calefactor inmerso dentro de la tina de acero inoxidable para mejor eficiencia en la transferencia del calor Todos los model...

Страница 2: ...ing Connector 2 30 3722 2 30 3722 2 30 3722 25 Control Temperatura Temperature Control 1 71 3134 1 71 3134 1 71 3134 26 Potenciometro Potentiometer 1 20 0260 1 20 0260 1 20 0260 27 Interruptor Piloto...

Страница 3: ...l cual detecta las fallas mas comunes en este tipo de equipos Cada vez que se enciende su Termoba o el sistema realiza una auto revision esta consiste en revisar los puntos cruciales para el funciona...

Страница 4: ...aparecer la temperatura de disparo de la alarma Para modificar este valor use nuevamente las teclas de incremento y de decremento el valor deber ser al menos un grado mayor que la temperatura de opera...

Страница 5: ...r no toque el fondo Espere a que se estabilice la temperatura y que el led de Ok est encendido compare las mediciones del display y de su medidor si existe alguna diferencia deber proceder en esemismo...

Страница 6: ...o tiene control sobre el uso y cuidado de estos elementos no seotorgagarant asobrelosmismos NOTAS IMPORTANTES NuncaoperesuTermoba osinaguasuficienteenlatina Retire elaguadelatinasino vaaoperarelTermob...

Страница 7: ...entilation in the area of the bottom For correct operation it is necessary that the user becomes familiarized with all the available controls and specifications shownineachmodel TANKFILLING OPERATIONT...

Страница 8: ...Set Point for two seconds and then theconfiguretemperature Withtheincreaseordecreasebuttonsyoucanmodifythisparametertothe desired operation temperature then press the menu button to continue with the...

Страница 9: ...27 with the menu button pressed turn on the equipment the display will show CAL and then a temperature adjust this number to the thermometer reading using the increase or decrease buttons and then pr...

Страница 10: ...cados por alteraciones hechas por personas no autorizadas por Felisa La responsabilidadmaximaenninguncasoseramayorqueelvalordelproductoinvolucrado Felisa se reserva el derecho de hace cambios o modifi...

Отзывы: