OBID
i-scan
®
Manual
ID ISC.PRHD102
FEIG ELECTRONIC GmbH
Page 75 of 125
H71101-2e-ID-B.doc
6.9. [0x74] Get Input
With this protocol the current status of the digital input IN1 (Switch) can be inquired at any time.
Host
→
Reader
1
2
3
4
5
6,7
0x02
7
COM-ADR
[0x74]
CRC16
Host
←
Reader
1
2
3
4
5
6
7
8,9
0x02
9
COM-ADR
[0x74]
STATUS
INPUTS
CRC16
INPUTS:
Bit:
7
6
5
4
3
2
1
0
Function:
-
-
-
-
-
-
-
IN1
b0
digital input = inactive
b1
digital input = active
6.10.
[0xA0] Reader-Login
The Reader-Login must be executed after every power up or
6.3. [0x63] RF Controller Reset
command,
if an access to the configuration parameters is desired.
Host
→
Reader:
1
2
3
4
5
6…9
10,11
0x02
11
COM-ADR
[0xA0]
READER-ID
CRC16
Host
←
Reader
1
2
3
4
5
6
7,8
6
8
COM-ADR
[0xA0]
STATUS
CRC16
READER-ID:
The READER-ID is a password which protects the configuration parameters from any
read and write access.
The READER-ID can be changed in the configuration block
NOTE:
•
A Reader-Login with wrong READER-ID cause a "Logout".
•
A "Logout" can be effected via the command
6.3. [0x63] RF Controller Reset
15
Содержание OBID i-scan ID ISC PRHD102
Страница 77: ...OBID i scan Manual ID ISC PRHD102 FEIG ELECTRONIC GmbH Page 77 of 125 H71101 2e ID B doc...
Страница 78: ...OBID i scan Manual ID ISC PRHD102 FEIG ELECTRONIC GmbH Page 78 of 125 H71101 2e ID B doc...
Страница 79: ...OBID i scan Manual ID ISC PRHD102 FEIG ELECTRONIC GmbH Page 79 of 125 H71101 2e ID B doc...
Страница 80: ...OBID i scan Manual ID ISC PRHD102 FEIG ELECTRONIC GmbH Page 80 of 125 H71101 2e ID B doc...