SGP10xxS Users Manual

                                     

 

Main technology indexes 

 

Signal Output function   

                                Output waveforms                          Sine wave, Square wave (Duty cycle 

adjustable) and Triangle wave (include 

sawtooth wave) 

                                Output amplitude                     

10Vp-p (signal output, no load)

MAX

 

                                                                                        About 0.1Vp-p (MIN) 

                                Output impedance                        51

Ω

±

10%(signal output) 

                                DC offset                                 

±3V(no load) 

                                Frequency range                      

0.01Hz ~ 2MHz(SGP1002S) 

0.01Hz ~ 5MHz(SGP1005S) 

0.01Hz ~ 8MHz(SGP1008S) 

                                Resolution                                

0.01Hz(10mHz) 

   

                      

Frequency accuracy                 

±

5

×

10

  

                            Frequency Stability                        

±

2

×

10

6

/

3 Hours

 

Sine wave distortion                      

0.8% (reference frequency is 1kHz)

 

Triangle linearity                     

98%    (0.01Hz~10kHz) 

Rise and fall time of square wave 

100ns 

Square Wave Duty range                1%~99%(digital control mode) 

 

 

TTL Output    function   

Frequency range                              0.01Hz ~ 2MHz (SGP1002S) 

0.01Hz ~ 5MHz (SGP1005S) 

0.01Hz ~ 8MHz (SGP1008S) 

                      Amplitude                                          >3Vp-p 

                      Fan Out                                              >20 TTL loads 

 

COUNTER    function   

                      Counter Range                                  0-4294967295 

                      Frequency Meter Range                    1Hz~60MHz 

                      Input Voltage Range                          0.5Vp-p~20Vp-p 

 

SWEEP    function   

                      Sweep mode                                      Linear sweep, Logarithmic sweep 

Frequency setting range                    0.01Hz ~ 2MHz (SGP1002S) 

0.01Hz ~ 5MHz (SGP1005S) 

0.01Hz ~ 8MHz (SGP1008S) 

Sweep range                                     

f

M1 

(pre-set) to f

M2 

(pre-set)

 

                      Sweep time                                        1s~99s 

 

Others   

                      Display                                        LCD1602 in English 

                      Save and Load Parameter            M0-M9(M0: default load) 

                      Size                                              135mm (Length)

×

65 mm(Width)

×

35mm(Height) 

                      Buzzer warning tone                    Can be turned On/Off by setting 

                      Production technology                Surface mounting technology, large scale integrated 

Содержание SGP10 S Series

Страница 1: ...SGP10xxS Users Manual SGP10xxS series Direct Digital Synthesis DDS Signal Generator Users Manual Rev1 1 2013 12 5 20140314 ...

Страница 2: ...ucture easy integration with customer products etc It s the ideal instrument for electronic enthusiasts laboratories production lines teaching and scientific researchers It can also be used as supporting module for industrial equipment Recently it has already widely used in teaching instruments and industrial control equipments It is a highly cost effective product It has excellent technical index...

Страница 3: ...me of square wave 100ns Square Wave Duty range 1 99 digital control mode TTL Output function Frequency range 0 01Hz 2MHz SGP1002S 0 01Hz 5MHz SGP1005S 0 01Hz 8MHz SGP1008S Amplitude 3Vp p Fan Out 20 TTL loads COUNTER function Counter Range 0 4294967295 Frequency Meter Range 1Hz 60MHz Input Voltage Range 0 5Vp p 20Vp p SWEEP function Sweep mode Linear sweep Logarithmic sweep Frequency setting range...

Страница 4: ...to switch between Frequency adjusting and Function adjusting The detailed condition frequency adjusting or function adjusting displayed after 2 As frequency adjusting pressing the button or to adjust position of cursor and pressing the button Confirm to toggle unit of frequency Hz kHz or MHz and then adjusting the Main knob to set the corresponding value of frequency Step frequency 1 kHz Step freq...

Страница 5: ...rm is TRGL 5 The DUTY means duty cycle It can be adjusted by rotating the Main knob SQUR adjusted from 1 to 99 while TRGL adjusted among 50 standard TRGL above 50 and below 50 both are different sawtooth waves SIN is disabled in this case WAVE SQUR WAVE SQUR WAVE TRGL WAVE TRGL WAVE TRGL 6 Press the button EXT FREQ COUNTER to switch between COUNTER function and EXT FREQ function 1 The COUNTER mean...

Страница 6: ... frequency is defined at M2 So if the sweep function need to be run the frequency values fM1 and fM2 must to be set firstly and make sure the values fM2 fM1 Choose storage position Save to 0 position is OK 8 The LOAD is function of loading the parameters of memory Operation is similar to SAVE 9 The TIME is the function of setting sweep time It can be set from 1 second to 99 seconds 10 The SWEEP is...

Страница 7: ... signal normally Safety Notes 1 Before using this instrument please check if the power supply is normal to ensure the normal use and personal safety 2 This instrument must be used in the technical index range 3 Please do not change the instrument circuit arbitrarily so as to avoid damaging equipment or endangering the safety Warning and personal injury Do not apply the product in the safety protec...

Отзывы: