RF emissions
CISPR 11

Harmonic emissions
IEC 61000-3-2

Voltage fluctuations/
flicker emissions
IEC 61000-3-3

Class B

Class A

Complies

This device is suitable for use in all establishments, including 
domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies buildings 
used for domestic purposes.

Guidance and Manufacturer's declaration – electromagnetic immunity

IEC

 

60601test

 

level

Compliance

 

level

Electromagnetic

 

environment

-

guidance

.

Electrostatic 
discharge (ESD) 
IEC 61000-4-2

±

6 kV contact

±

8 kV air

±

6 kV contact

±

8 kVair

Floors should be wood, concrete or 

ceramic tile. If floors are covered 

with synthetic material, the relative 

humidity should be at least 30%.

The device is intended for use in the electromagnetic environment specified below. 
The  customer or the user of the device should assure that it is used in such an environment

Immunity

 

test

Electrical fast 
transient/burst 
IEC 61000-4-4

±

2 kV for power supply lines

±

1 kV for input/output lines

Mains power quality should be that of 

a typical commercial or hospital 

environment.

Mains power quality should be that 

of a typical commercial or hospital 

environment.

Surge IEC 
61000-4-5

±

1 kV line to line

±

2 kV line to earth

Voltage dips,
short interruptions
and voltage
variations
on power supply
IEC 61000-4-11

Mains power quality should be that 
of a typical commercial and/or 
hospital environment. If the user of 
this device requires continued 
operation during power mains 
interruptions, it is recommended that
the device be powered from an 
uninterruptible power supply or battery.

18

<5 % UT(>95 % dip in UT) for
0.5 cycle

40 % UT(60 % dip in UT) for 
5 cycles

70 % UT(30 % dip in UT) for
25 cycles

<5 % UT(>95 % dip
 in UT) for 0.5 cycle

40 % UT(60 % dip in
 UT) for 5 cycles

70 % UT(30 % dip in 
UT) for 25 cycles

±

1 kV line to line

±

2 kV line to earth

±

2 kV for power 

supply lines*1)

<5 % UT(>95 % dip in UT) for
5 sec

<5 % UT(>95 %
dip in UT)for 5 sec

Содержание Air Pro

Страница 1: ...Version V2 Air Pro 120mm 100mm H 100g Model No Air Pro User Manual for portable Mesh Nebulizer...

Страница 2: ...user Installation 4 How to use 5 Cleaing and Disinfection 6 Storage and Maintenance 7 Contraindications Precautions Notice and Warning 8 Trouble Shooting Tips 9 After sales Service 10 Symbol Descripti...

Страница 3: ...urpose Clean and disinfect the medication cup and accessories before using or not using the unit for quite a while Please stop using the device if the components are damaged or fall into the water acc...

Страница 4: ...zed particles through the suction mask or nozzle to guide the patient s respiratory system in order to achieve the purpose of inhalation therapy Respiratory system is an open system the liquid is atom...

Страница 5: ...ological salineunder conditions of a temperature of 25 3 C and a humidity of 60 5 R H And theequivalent particle size distribution curve of the fog particles measured under theseconditions is as follo...

Страница 6: ...F C E A B D A Medication Cover B Medication Cup C Spray Connector D Main Unit E ON OFF Switch F Slide G PRESS Switch H Micro USB Cable 3 Installation and User Instructions 1 1 The nebulizer has a USB...

Страница 7: ...to the local environmental regulation Do not dispose together with the rubbish to avoid environment pollution Do not dismantle or repair the equipment or components do not dismantle replace the batter...

Страница 8: ...following Figure 1 Note Follow the doctor s advice to inject the liquid do not exceed the maximum capacity of the medication cup the maximum capacity can be seen on medication cup scale After injectin...

Страница 9: ...ing a direct inhalation b by mouthpiece c By mask 9 Note Clean disinfect dry all parts before installation b Connect the mask a Connect the mouthpiece 3 When finishing nebulization press ON OFF switch...

Страница 10: ...ection 4 Liquid will be coagulated around the spray connector and mesh disc which will affect the nebulization effect after nebulization It should stop nebulization remove the mask mouthpiece and othe...

Страница 11: ...cover mask mouthpiece etc should be disinfected after use Methods are as below 2 Disinfection 11 Note Please avoid cleaning the main unit in water in case water entering Please keep the main unit and...

Страница 12: ...ndition b The nebulizer should be promptly cleaned and disinfected after use the medication cup and other accessories should store into the packing box after completely drying Stored in the required e...

Страница 13: ...ipment personally Please clean and disinfect the unit after use and refer to the clean and disinfectcharpter The nebulizer is for medication atomization The nebulizer is for medication atomization The...

Страница 14: ...t5 0V 1 0A Do not touch the center of spray mesh by hand or other sharp objects May cause damage and can not be used The nebulizer has no moisture and dust function and the product should not be store...

Страница 15: ...pray mesh 4 The spray mesh maybe blocked you can drop 2 or 3 drops of white vinegar into the medication cup with 3 6ml water then fully nebuilzed Clean the medication cup todisinfection for next nebul...

Страница 16: ...pretation of the warranty card which may be subject to change without prior notice 1 The device is warranted to be free from defects materials and workmanship appearing within 1year from the date of p...

Страница 17: ...nvironment Group 1 17 Note Warning refer to enclosed file Upward Avoid Moisture The application part of Type BF No toxic and harmful substances and elements Refer to instruction manual booklet NOTE On...

Страница 18: ...it is used in such an environment Immunity test Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical co...

Страница 19: ...n the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic pro...

Страница 20: ...ter in watts W according to the transmitter manufacturer 0 01 0 1 1 10 100 0 12 0 38 1 2 3 8 12 0 12 0 38 1 2 3 8 12 0 23 0 73 2 3 7 3 23 Precautions 1 Equipment or systems should not be used or stack...

Страница 21: ...6 6 Serial No Serial No Email info feellife com Serial No Serial No Email info feellife com 21...

Страница 22: ...Feellife Medical Inc Product Portable mesh nebulizer Model Air Pro Inspection date Inspector CERTIFICATE OF INSPECTION 22...

Страница 23: ...thpiece User manual Warranty card Certificate of inspection Micro USB cable Quantity If Included Yes No Configuration list User Guidance Item Unit Medication cup Adult mask Velvet bag 23 Warning badge...

Страница 24: ...755 2699 3192 After sales service FEELLIFE MEDICAL INC Service life 5 years Website http www feellife com Room 1903 Building A No 9 Furong Road Tantou Community Songgang Subdistrict Bao an District S...

Отзывы: