background image

12

If the power to the solenoid valve coil
is correct, the following symptoms
indicate that cleaning the solenoid
vavle is required:
• Sluggish valve operation

• Excessive noise

• Constant water spray

• Streaming water 

To clean the solenoid valve:

1.  Turn the power to the furnace OFF.

2.  Turn the humidistat to the OFF

position

3.  Turn off the water supply at the

saddle valve

4.  Disconnect the power and water

lines from the humidifier

5.  Remove the humidifier from the

duct.

6.  Remove the drip sleeve from the

nozzle assembly.

7.  Using a 3/4” wrench, remove the

nozzle adapter.

8.  Remove the solenoid valve

assembly from the case using a
Phillips head screwdriver.

9.  Remove the brass nipple from the

OUT port of the valve body.
Remove the street elbow and brass
adapter from the IN port of the
valve.

10.  Remove the filter from inside the

inlet brass adapter using a small
nail or paperclip. 

Note: Be careful not to
puncture the filter.

11.  Remove the coil from the assembly

by turning the valve body counter
clockwise.

12.  Clean and remove all particulates

from the brass fittings, filter,
plunger/spring assembly and
orifices in the valve body using a
50/50 solution of white vinegar
and water or a liquid humidifier
cleaner. 

13.  Reassemble the parts, paying

careful attention to the diagram
below for the correct placement.

14. After completing the assembly,

operate the unit to be sure it is
functioning properly.

Maintenance

NOTE: Teflon tape may be used on the threads for
re-assembly, however, do not use pipe dope, as it
will clog the nozzle assembly.

Coil

IN

OUT

O-Ring
(Rubber)

Valve

Body

(Brass)

Brass Nipple (To Nozzle Adapter)

Street 
Elbow

Brass 
Adapter

Filter

To Water Tubing

Plunger/Spring
Assembly

Содержание Herrmidifier Mister 50

Страница 1: ...To Check for Leaks 10 To Clean Spray Nozzle 11 To Clean Solenoid Valve 12 Unit Diagram and Parts List 13 Warranty Back cover Manual for Installation Operation Maintenance Duct Mount Atomizing Humidifi...

Страница 2: ...y are many They include both personal comfort as well as the preservation of furniture draperies carpets wooden floors and cabinets paintings pianos etc Your home will be more comfortable at a lower t...

Страница 3: ...me of 70 This chart uses A R I standard designations A Tight Home is assumed to be well insulated with vapor barriers tight storm windows and doors and a dampered fireplace Air exchange rate of 5 chan...

Страница 4: ...185 4 44 225 5 44 50 250 6 00 300 7 20 75 375 9 00 445 10 68 1 0 500 12 00 600 14 40 WATER PRESSURE Oil Nozzle 80 PSI 100 PSI size GPH GPD GPH GPD 37 275 6 60 300 7 20 50 350 8 40 400 9 60 75 550 13...

Страница 5: ...hermostat can be located upstream from the nozzle For plenum mounting the nozzle should always be pointed in the up position When the position of the nozzle has been determined tighten the nozzle sole...

Страница 6: ...ourself 8 Remove the Retainer and 4 sheet metal screws from the parts bag 9 Carefully place the Retainer inside the duct and fasten loosely with 1 screw 10 Place the bottom lip of the humidifier on th...

Страница 7: ...the end of the tubing in hot water to soften the plastic Place the insert against a flat surface and press the tubing onto the insert 20 Fasten tubing assembly securely to the saddle valve but do not...

Страница 8: ...hermostat Mister 50 24v Humidifier Solenoid Valve Electrical connections 24v Fan Limit Fan Motor Humidistat Thermostat Mister 50 1 120v Humidifier Solenoid Valve Electrical connections 120v WARNING Im...

Страница 9: ...on BEFORE the humidifier 3 The solenoid valve will make an audible click when it operates either on or off 4 If necessary adjust the furnace fan control to a lower setting to allow the blower to turn...

Страница 10: ...on The humidifier should activate after the furnace blower comes on and shut off before the furnace blower turns off If this does not occur readjust the furnace fan switch as required If applicable se...

Страница 11: ...er is not the source of these impurities Maintenance To check for leaks 1 Turn the furnace and humidistat on and ensure that the humidifier has operated for at least 1 minute 2 Remove the screws from...

Страница 12: ...ench into the set screw and turn it counter clockwise and remove the components 10 Rinse and clean all parts Suggested cleaners include 50 50 solution of white vinegar and water or liquid humidifier c...

Страница 13: ...brass adapter from the IN port of the valve 10 Remove the filter from inside the inlet brass adapter using a small nail or paperclip Note Be careful not to puncture the filter 11 Remove the coil from...

Страница 14: ...13 Unit Diagram 9 8 13 10 11 12 20 Humidifier 20 19 17 19 15 18...

Страница 15: ...olt Valve Assembly 12004 002 10 Street Elbow 12009 12009 11 Adapter Brass G 157 G 157 12 Filter Conical G 125 G 125 13 Cover 12002 12002 14 Humidistat 24 volt not shown 352680 004C 14a Humidistat 120...

Страница 16: ...edders com Humidifiers which have been installed or become part of real estate cannot be returned CareCo will receive freight prepaid only removable parts or components of such defective humidifiers 6...

Отзывы: