Guidance and manufacture’s declaration – electromagnetic immunity

 

The   

[EQUIPMENT or SYSTEM] i

s

 intended for use in the electromagnetic environment specified below. The customer 

or the user of   

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment.

 

Immunity test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment - 

guidance 

Conducted RF 
IEC 61000-4-6 

3 V

rms

 

150 kHz to 80 MHz 

Not   

applicable

 

Portable  and  mobile  RF  communications 
equipment  should  be  used  no  closer  to  any 
part  of the 

[EQUIPMENT or  SYSTEM]

,  including 

cables,  than  the  recommended  separation 
distance 

calculated 

from 

the 

equation 

applicable to the frequency of the transmitter. 

Recommended separation distance 
 

P

d

167

.

1

=

          

 

P

d

167

.

1

=

         80 MHz to 800 MHz 

 

P

d

333

.

2

=

         800 MHz to 2.5 GHz 

Where 

P

 is the maximum output power rating 

of the transmitter in watts (W) according to the 
transmitter  manufacturer  and 

d

  is  the 

recommended  separation  distance  in  metres 
(m). 
Field  strengths  from fixed  RF  transmitters,  as 
determined by an electromagnetic site survey,

a

 

should  be  less  than  the  compliance  level  in 
each frequency range.

b

 

Interference  may  occur  in  the  vicinity  of 
equipment marked with the following symbol: 

 

 

 

 

Radiated RF 

IEC 61000-4-3 

 
 
 
 
 

3 V/m 

80 MHz to 2.5 GHz 

 

3 V/m 

NOTE 1        At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2        These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

a

         Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey  should  be  considered.  If  the  measured  field  strength  in  the  location  in  which  the   

[EQUIPMENT  or 

SYSTEM] 

is used exceeds the applicable RF compliance level above, the   

[EQUIPMENT or SYSTEM] 

should be 

observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may  be 
necessary, such as reorienting or relocating the   

[EQUIPMENT or SYSTEM]

b

        Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

 

 

 

 

 

 

 

 

Содержание FT-C21Y

Страница 1: ...the manual appropriately for future reference This User Manual is suitable to the following models FT C21Y FT C22Y FT C23Y FT C24Y FTC11B FT C12B FT C21Y V FT C22Y V FTC23Y V FT C24Y V FT C11B V FT C...

Страница 2: ...ual by using a non invasive technique in which an inflatable cuff is wrapped around the arm The cuff circumference is limited to 8 7 inches to 12 6 inches Specification Product Name Arm Automatic Bloo...

Страница 3: ...by internal electromagnetic valve Display LCD digital display It can show Pressure Pulse Date Time Color Backlight display on LCD Optional White backlight display when power on Green backlight display...

Страница 4: ...o protect natural environment by respecting national and or local recycling regulations when disposing of the battery and the product at the end of their useful live II PRECAUTION FOR USE AND MAINTENA...

Страница 5: ...DO NOT wrap the cuff on the arm on the side of a mastectomy 13 DO NOT use the device in the vicinity of flammable gases such as those used for anesthesia Doing so could ignite the gases and cause an e...

Страница 6: ...DO NOT attempt to disassemble the device User can open battery cover for new battery installation DO NOT bend the cuff or air hose excessively 5 NEVER clean the blood pressure monitor with thinners or...

Страница 7: ...n Deflation Indicator 8 Blood Pressure Measurement Unit 9 Battery Symbol 10 Irregular heartbeat Indicator 11 Memory Record Number Back light display function is optional IV Measure Procedure Battery L...

Страница 8: ...ced Figure 3 Figure 4 NOTE AC adapter is optional for power supply Only use the adapter provided by manufacturer that is complying with EC60601 1 standard requirements Do not touch the AC adapter with...

Страница 9: ...or Kpa and the mmHg is the definition unit When the machine is turned off press the button ON OFF more than about 10 seconds till LCD blinks then press Memory button to switch between mmHg and Kpa Arm...

Страница 10: ...S CUFF PRESSURE MAY EFFECT BLOOD FLOW AND CAUSE HARMFUL INJURY Figure 6 Note The cuff is TYPE BF APPLIED PART Measuring Process POSTURE FOR TAKING BLOOD PRESSURE MONITOR Cuff should be at the same hei...

Страница 11: ...will speak out the displayed blood pressure heart rate If an irregular heartbeat is detected the IRREGULAR HEARTBEAT symbol will appear and blink in the display screen See Figure9 NOTE Do not self di...

Страница 12: ...This model has 2 Person memory banks and 90 memories storage for each Press and release SET button can prompt to P1 Person 1 or P2 Person 2 as your ID to access the measurement for the first time oper...

Страница 13: ...85 89 Stage 1 Hypertension 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non...

Страница 14: ...each measurement finish The device will speak out the last time displayed memory blood pressure systolic and diastolic pressure and heart rate when reading memory result Note The voice function is ONL...

Страница 15: ...tion method After testing the result displayed on LCD including systolic pressure diastolic pressure and pulse will be automatically sent to the blue tooth receiving device such as mobile phone The ad...

Страница 16: ...during the day and will also be affected by emotional and physical condition LCD shows Er U Insufficient inflation Measure again LCD shows Er H Inflation over 305 mmHg Measure again LCD shows Er 1 Un...

Страница 17: ...eception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receivi...

Страница 18: ...ns power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not applicable Mains power quality should be that of a typic...

Страница 19: ...urvey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz t...

Страница 20: ...117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation di...

Страница 21: ...PART Symbol for Follow operating instructions IP21 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...

Отзывы: