Arm Automatic Blood Pressure Monitor 

(Electronic Sphygmomanometer) 

User Manual

 

 

  Thank your for purchasing our product.   

  Please read this manual carefully before using the product.   

  Please keep the manual appropriately for future reference.   

  This User Manual is suitable to the following models: FT-C21Y, FT-C22Y, FT-C23Y, 

FT-C24Y, FTC11B, FT-C12B,   FT-C21Y-V, FT-C22Y-V, FTC23Y-V, FT-C24Y-V, 

FT-C11B-V, FT-C12B-V, FT-C11B-UR and FT-C11B-BT. 

 

 

 

 

Table of Contents 

 

(I) General Information 

(II) Precaution for Use and Maintenance 

(III) Name of Each Part 

(IV) Measure Procedure 

(V) Troubleshooting 

(VI) Model List 

Содержание FT-C21Y

Страница 1: ...the manual appropriately for future reference This User Manual is suitable to the following models FT C21Y FT C22Y FT C23Y FT C24Y FTC11B FT C12B FT C21Y V FT C22Y V FTC23Y V FT C24Y V FT C11B V FT C...

Страница 2: ...ual by using a non invasive technique in which an inflatable cuff is wrapped around the arm The cuff circumference is limited to 8 7 inches to 12 6 inches Specification Product Name Arm Automatic Bloo...

Страница 3: ...by internal electromagnetic valve Display LCD digital display It can show Pressure Pulse Date Time Color Backlight display on LCD Optional White backlight display when power on Green backlight display...

Страница 4: ...o protect natural environment by respecting national and or local recycling regulations when disposing of the battery and the product at the end of their useful live II PRECAUTION FOR USE AND MAINTENA...

Страница 5: ...DO NOT wrap the cuff on the arm on the side of a mastectomy 13 DO NOT use the device in the vicinity of flammable gases such as those used for anesthesia Doing so could ignite the gases and cause an e...

Страница 6: ...DO NOT attempt to disassemble the device User can open battery cover for new battery installation DO NOT bend the cuff or air hose excessively 5 NEVER clean the blood pressure monitor with thinners or...

Страница 7: ...n Deflation Indicator 8 Blood Pressure Measurement Unit 9 Battery Symbol 10 Irregular heartbeat Indicator 11 Memory Record Number Back light display function is optional IV Measure Procedure Battery L...

Страница 8: ...ced Figure 3 Figure 4 NOTE AC adapter is optional for power supply Only use the adapter provided by manufacturer that is complying with EC60601 1 standard requirements Do not touch the AC adapter with...

Страница 9: ...or Kpa and the mmHg is the definition unit When the machine is turned off press the button ON OFF more than about 10 seconds till LCD blinks then press Memory button to switch between mmHg and Kpa Arm...

Страница 10: ...S CUFF PRESSURE MAY EFFECT BLOOD FLOW AND CAUSE HARMFUL INJURY Figure 6 Note The cuff is TYPE BF APPLIED PART Measuring Process POSTURE FOR TAKING BLOOD PRESSURE MONITOR Cuff should be at the same hei...

Страница 11: ...will speak out the displayed blood pressure heart rate If an irregular heartbeat is detected the IRREGULAR HEARTBEAT symbol will appear and blink in the display screen See Figure9 NOTE Do not self di...

Страница 12: ...This model has 2 Person memory banks and 90 memories storage for each Press and release SET button can prompt to P1 Person 1 or P2 Person 2 as your ID to access the measurement for the first time oper...

Страница 13: ...85 89 Stage 1 Hypertension 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non...

Страница 14: ...each measurement finish The device will speak out the last time displayed memory blood pressure systolic and diastolic pressure and heart rate when reading memory result Note The voice function is ONL...

Страница 15: ...tion method After testing the result displayed on LCD including systolic pressure diastolic pressure and pulse will be automatically sent to the blue tooth receiving device such as mobile phone The ad...

Страница 16: ...during the day and will also be affected by emotional and physical condition LCD shows Er U Insufficient inflation Measure again LCD shows Er H Inflation over 305 mmHg Measure again LCD shows Er 1 Un...

Страница 17: ...eception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receivi...

Страница 18: ...ns power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not applicable Mains power quality should be that of a typic...

Страница 19: ...urvey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz t...

Страница 20: ...117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation di...

Страница 21: ...PART Symbol for Follow operating instructions IP21 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...