![Faxitron Bioptics TruNode S10 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/faxitron-bioptics/trunode-s10/trunode-s10_user-manual_546853023.webp)
TruNode
®
Gamma Probe System User Manual
23
stacked with other equipment. If adjacent or stacked use is necessary,
the TruNode® Gamma Probe System should be observed to verify
normal operation in the configuration in which it will be used.
The Trunode® Gamma Probe System transmitter and receiver use
Bluetooth
® SMART
technology, with a frequency band of 2.400 GHz-
2.4835 GHz and GFSK modulation. The effective radiated power of the
transmitter is 1.2mW.
The Trunode® Gamma Probe System may be interfered with by other
equipment, even if that other equipment complies with CISPR
EMISSION requirements.
Guidance and declaration – electromagnetic emissions
The TruNode® Gamma Probe System is intended for use in the electromagnetic environment
specified below. The customer or the user of the TruNode® Gamma Probe System should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions CISPR 11
Group 1
The TruNode® Gamma Probe System
uses RF energy only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The TruNode® Gamma Probe System is
suitable for use in all establishments
other than domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies