CALIBRATION
Regular calibration of the probes is important.
The calibration is carried out by means
of the reference solution supplied.
Clean the SALINITY MANAGER with the
Cleaning Solution for about 2 minutes.
Then clean the SALINITY MANAGER in
osmosis water and shake any residual water
out of the device.
Unscrew the upper lid and immerse the
SALINITY MANAGER in the calibration solu-
tion. If necessary, adjust the measured value
to
35 ppt
with the included screwdriver.
Further technical information and operating
aids can be found in our
HTU ( How To Use ) SALINITY MANAGER
For download at
www.faunamarin.de DownloadCenter
EN
13
ON OFF
ON OFF
Cleaning
Solution
(2 min)
osmosis
water
ON OFF
1.
ON OFF
35 ppt
✓
35 .0
SALINITY MANAGER
calibration solution
2.