Functional
Description
ComTec GmbH
3-3
3.3.4. Adjust
ULD
Up/Down buttons set the upper level discriminator, continuously adjustable from LLD threshold to
+10 V.
3.3.5. Adjust Coincidence / Anti Coincidence
Up/Down buttons set the coincidence/anti-coincidence GATE mode for ADC and SVA modes.
3.3.6. Normal
operation
When the device is in normal operation mode the UP/DOWN buttons have two different functions.
In PHA/SVA mode the Up/Down buttons select the conversion range of the ADC. The conversion
range of each channel cycles through 8k - 4k - 2k - 1k - 512 - 256 - 8k - etc.
In TDC mode the UP/DOWN buttons select the time conversion range. It cycles through 50 ns -
100 ns - 200 ns - 500 ns - 1 us - 2 us - 5 us - 10 us - 20 us - 50 ns - etc.
3.3.7. Factory
reset
To reset all device parameters to the default values as the 7072T was shipped from the factory
press both DOWN BUTTONS simultanously during power-up.