FaroArm User Guide
January 2007
37
Chapter 2: CAM2 Measure Devices Menu
Single Hole Method
The Single Hole calibration is
performed using the FARO probe
calibration cone or a 5mm machine
drilled hole. The hole does not have to
be exactly 5mm, but it must be smaller
than the probe’s diameter with a smooth
seat. If the G
UIDANCE
check box is
clear for the current probe, use the
following steps.
All of the points in this method will be
collected by holding down the FRONT
button. The FaroArm will collect points as fast as possible ("scanning")
until you release the FRONT button.
1
Place the ball probe in the hole. Start in a vertical position.
2
Press and hold the FRONT button. Sweep the handle down to
to one of the horizontal positions. Be sure that the ball probe
remains seated in the hole.
3
Release the FRONT button and repeat for each position.
•
Digitize points in the hole and sweep to
position #1
Position #1
Position #2
Position #3
08M46E00_FaroArmUSB.book Page 37 Monday, January 15, 2007 1:12 PM
Содержание FaroArm P08
Страница 1: ...FAROARM USER GUIDE JANUARY 2007 08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...
Страница 2: ...08M46E00_FaroArmUSB book Page 2 Monday January 15 2007 1 12 PM...
Страница 4: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 48: ...08M46E00_FaroArmUSB book Page 42 Monday January 15 2007 1 12 PM...
Страница 54: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 64: ...08M46E00_FaroArmUSB book Page 10 Monday January 15 2007 1 12 PM...
Страница 70: ...08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 76: ...D 6 08M46E00_FaroArmUSB book Page 6 Monday January 15 2007 1 12 PM...
Страница 80: ...08M46E00_FaroArmUSB book Page 4 Monday January 15 2007 1 12 PM...
Страница 81: ...08M46E00_FaroArmUSB book Page 1 Monday January 15 2007 1 12 PM...