Faro EVA Скачать руководство пользователя страница 11

EVA           

 

USER MANUAL - EN 

Edition 1.0 April 2020    

 

 

 

 

 

 

                                      Pag. 11 di 14 

              

6

 

CLEANING AND DISINFECTION 

  

 

Warning against danger of wear and corrosion and falling suspended mass 

For all metal or plastic parts it is strictly forbidden to use substances that are  

-

 

abrasive,  

-

 

corrosive,  

-

 

acids,  

-

 

substances containing chlorine or chloride ions, phosphorous or phosphorous ions,  

-

 

detergents with Trilene base, petrol, white spirit, chlorine or similar. 

Do not use detergents-disinfectants containing the following substances to clean plastic parts: 

-

 

Ammonium Hydroxide 

-

 

Sodium Hydroxide 

-

 

Hydrogen peroxide 

-

 

Ammonium Chloride 

-

 

Methylene chloride 

-

 

Methyl alcohol 

-

 

Acids and corrosive substances of all kinds. 

It is forbidden to directly spray any chemical substance on the device. 

It is forbidden the use of wet wipes without rinsing.

 

   

Faro tested and suggests the use of the following disinfectants, for plastic parts and metal parts: 
-

 

Durr FD366 Sensitive 

-

 

Faro Perflex Advanced 

-

 

Water-alcohol based disinfectants with 70% isopropyl alcohol or ethanol are suitable. 

 
6.1

 

CLEANING OF THE REFLECTING PARABOLAS 

Cleaning must be carried out using a soft cloth in cotton or absorbent cotton with ethyl alcohol.  
Water-alcohol based disinfectants are suitable with 70% isopropyl alcohol or ethanol.  

 

 

Caution - potential damage or wear on the parabolas

 

Never spray detergent directly on the parabolas. 
Cleaning operations on the parabolas must be carried out wearing gloves, to avoid leaving fingerprints on the surfaces. 

Never use detergents containing surfactants or water-repellents that depositing can leave streaks. Slight streaking will not 

prejudice the quality of the light. 
Products differing from those suggested could damage the parabolas. 

If in doubt, contact FARO customer care.

 

 

6.2

 

CLEANING AND DISINFECTION OF THE HEAD 

Cleaning must be carried out using a soft cloth in cotton wetted with disinfectant solution. 
Always squeeze the cloth to remove all the liquid in excess.  

 

6.3

 

CLEANING AND DISINFECTION OF ARMS 

Always use a cloth soaked in disinfectant approved to disinfect the surfaces and pass it over.  

Always squeeze the cloth to remove all the liquid in excess. 
 

7

 

STERILIZATION OF THE HANDLES 

 

Warning - danger of cross contamination 

The handles are not supplied sterile, they must therefore be sterilised before 
use.  The handles must be sterilised before each patient. 

 
7.1

 

REMOVAL OF THE HANDLES 

To remove the handle, unscrew knob “A” and remove it from the support. 

 

7.2

 

DECONTAMINATION AND DISINFECTION 

Before sterilising the handles, they must be decontaminated and disinfected.

 

To disinfect, Faro has tested the following products for disinfection:  

 

Faro Perflex Advance 

 

Durr FD366 Sensitive 

 

  

 

WARNING - danger of plastic breaking 

The handles cannot be disinfected by thermo-disinfection. 

 

7.3

 

STERILIZATION 

The handles must be packaged in compliance with EN 868-5. 
The  handles  can  be  sterilised  with  standard  cycles  121°/134°  C  up  to  two  hundred  (200)  cycles  or  however  up  to  loss  of  the 
mechanical performance. 
The parameters of the sterilisation cycle are as follows: 

 

Cycle EN 13060 

Temperature 

Pressure 

Holding Time Minimum 

121°C 

207 kPa 

15 min 

134°C 

308 kPa 

3 min 

 
 
 

Содержание EVA

Страница 1: ......

Страница 2: ...LE WHITE VERSION 7 4 4 COMPOSAVE SETTING ON TUNABLE WHITE VERSION 8 4 5 MINIMUM INTENSITY SETTING ON SOLAR VERSION 8 4 6 AUTO ON SETTING 8 4 7 SWITCHING ON OFF THEIA 8 4 8 REMOTE CONTROL 8 4 9 SYNCRO...

Страница 3: ...nd the competent authority of the Member State in which the user and or patient is established 1 SYMBOLS USED 1 1 SYMBOLS USED IN THIS MANUAL WARNING The paragraphs marked with this symbol contain ins...

Страница 4: ...ce and is intended for illuminating the oral cavity and oral structures of patients in dentistry In the normal use the device is positioned distance of 700mm from the operative area the distance for w...

Страница 5: ...DIGIT Fixed Arm 4 DIGIT Light Source Integrated Camera 5 DIGIT Command radiofrequency RF 6 DIGIT Cables 7 DIGIT Custom 1 8 9 DIGIT 5 U 0 Only Head No arms 0 Only Head No arms 0 Tunable White 0 Joystic...

Страница 6: ...eters are memorized by the lamp and made available at every switching on Last light intensity setting light colour temperature Setting for Tunable White variant Warning Do not use the device if parts...

Страница 7: ...or light changes according to the illuminance level according to the images below Light Intensity Minimum Light Intensity Medium Minimum Light Intensity Medium Maximum Light Intensity Maximum 4 3 CHAN...

Страница 8: ...hen the on mode in set in on position the lamps turn on automatically without a specific command from the user when there is power supply 4 7 SWITCHING ON OFF THEIA The light on the fixed arm secondar...

Страница 9: ...Light it is necessary to repeat the procedure from point 1 3 After the blue LED switches on the room light there are 60 seconds to confirm the Pairing by pressing the programming button placed on the...

Страница 10: ...arter 1 2 must be well secured and safety screws intact The screws under carter 3 must be well secured Yearly Carter 3 Carter 2 Carter 1 Service Engineer Check the absence of any oxidation into joints...

Страница 11: ...es Never use detergents containing surfactants or water repellents that depositing can leave streaks Slight streaking will not prejudice the quality of the light Products differing from those suggeste...

Страница 12: ...0 60 Hz 32 V dc 10 Max Power 24 V ac 26 VA 32 V dc 14 VA 24 V ac 40 VA 32 V dc 28 VA Recommended fuses for installation not supplied in charge to installation 24 V ac T1 6AL 250V 32 V dc T1 25AL 250V...

Страница 13: ...IONS The device in the original packaging can be transported and stored for a maximum period of 15 weeks if the following environmental conditions are met Environmental temperature from 20 C to 70 C R...

Страница 14: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...