Cable Requirements for EMC Compliance
Select and install cables for the NI-9229 with BNC in accordance with the following
requirements:
■
Install a clamp-on ferrite bead (NI part number 782801-01) on the BNC cable
for each channel that you are connecting to on the NI-9229 with BNC.
■
Clamp-on ferrites must be connected to the BNC cable as close to the
module as possible with a full turn as shown in the following figure. Placing
the ferrite elsewhere on the cable noticeably impairs its effectiveness.
Figure 1. Installing a Ferrite
Special Conditions for Marine Applications
Some products are approved for marine (shipboard) applications. To verify marine
approval certification for a product, visit
, search by
model number, and click the appropriate link.
Notice
In order to meet the EMC requirements for marine applications,
install the product in a shielded enclosure with shielded and/or filtered
power and input/output ports. In addition, take precautions when
designing, selecting, and installing measurement probes and cables to
ensure that the desired EMC performance is attained.
Preparing the Environment
Ensure that the environment in which you are using the NI-9229 meets the following
specifications.
Operating temperature (IEC 60068-2-1, IEC 60068-2-2)
-40 °C to 70 °C
ni.com