Please note! It is a good practice to erase
the average memory after completing a
measuring session. This prevents any old
values affecting later measurements.
The avg. memory blinks If it is full.
3.2. Adjusting the calibration
If you have had a sample tested in an oven and
the moisture is different to what the Wile 27
reads. Then you can adjust your Wile 27 to
match the oven test, and this adjustment can
remain in the settings to change future
measurements.
The calibration can be ad/- 10 moisture
per cent
3.2.1. If you want to add the result
Press
MENU
twice when the result is on the
display. Three bars will appear on the top of the
display and after that the moisture reading. Now
add 0,1 percent to the result with each press of
MENU
.
3.2.2. If you want to deduct from the result
Press
MENU
three times when the result is on
the display. Three bars will appear on the
bottom of the display and after that the moisture
reading. Now deduct 0,1 percent from the result
with each press of
MENU.
Please note!
The adjustment of the calibration
is specific for each of the 1 to 6 scales.
3.2.3. Adjustment of the calibration
An adjustment to the calibration will appear on
the display after the next
run
.
E.g. "
-.5
". - 0,5 per cent units have been
deducted from the result before displaying it.