SECTION 2
SAFETY CONSIDERATIONS
Safe operation of the Differential Sticking Tester requires that the operator understand and practice the
correct assembly and operation of the equipment. Improper assembly, operation, or the use of defective
parts poses the possibility of cell leakage or failure which could result in serious injury and damage.
Following is a list of suggestions that should be observed to assure safe operation and maintenance of
the Differential Sticking Tester.
A. Safe Pressurization
1. Always use either Carbon Dioxide or Nitrogen. Never connect the Differential Sticking Tester
to Compressed Air, Oxygen or other non-recommended gas. If Nitrogen is used it must be
supplied in an approved Nitrogen gas cylinder or the nitrogen supply system must be built into
the laboratory. Nitrogen Cylinders must be secured to meet safety standards. Carbon
Dioxide is normally supplied in small cartridges which contain about 900 psi (6206 kPa)
pressure. They are primarily used for field operations. Do not allow these cartridges to be
heated or exposed to fire. They can explode if overheated.
2. Maintain pressure regulators in good condition. Never use oil on pressure regulators. Leaking
pressurization systems should be repaired or replaced. Gauges, fittings and hoses should be
kept in good condition and leaks should be found and corrected. If Nitrogen is being used,
periodically test the safety relief valve to verify they will relieve if excessive pressure should
occur. Never plug or bypass the safety valve.
3. Before pressurizing the Differential Sticking Tester always check to be sure the regulator tee
screw is backed out until free turning thereby putting the regulator in the closed position.
Insert and puncture the CO
2
cartridge or open the nitrogen gas supply valve, then adjust the
regulator. Do not attempt to pressurize higher than the equipment is rated. When de-
pressurizing, shut off the supply pressure, bleed the system of pressure, and then back out
the regulator Tee screw.
B. Safe Test Cell Maintenance
The Differential Sticking Tester Cell Assembly constitutes a PRESSURE VESSEL.
The Safety Precautions listed should be followed to assure safe operation.
1. Cell material should be compatible with the test sample.
2. Cells that show signs of stress cracking, severe pitting, or have damaged threads must not be
used.
C. Safe Operation
The Differential Sticking Tester is a relatively small and light weight instrument. Observe the following
cautions in its operation to prevent injury and damage due to the instrument slipping on the work
bench or accidentally being knocked off the bench.
1. Before attempting to tighten Cell Cap onto Cell Body, Refer to Fig. 3, make sure Bolt [26], Fig. 2 is
installed and tightened. This will insure the Cell does not slip off the Stage.
Содержание 21150
Страница 1: ...Model 21150 Differential Sticking Tester Instruction Manual Instruction Manual Part No 206907 Rev D ...
Страница 4: ...iv ...
Страница 6: ...2 ...
Страница 9: ...5 ...
Страница 10: ...6 ...
Страница 13: ...9 ...
Страница 14: ...10 ...
Страница 16: ...12 ...
Страница 22: ...18 ...
Страница 23: ...19 Fig 7 Exploded View Pressure Regulator ...
Страница 24: ...20 ...
Страница 28: ...24 ...
Страница 31: ...27 ...