Guidance and manufacturer´s declaration – electromagnetic emission
The FDTH3400 is intended for use in the electromagnetic environment specified below. The customer or the user of
FDTH3400 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The FDTH3400 uses RF energy only for its internal function. There for,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The FDTH3400 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
N/A
Guidance and manufacturer's declaration – electromagnetic immunity
– for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The FDTH3400 is intended for use in the electromagnetic environment specified below. The customer or the user of
the FDTH3400 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ±
8 kV,± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ±
8 kV,± 15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %. If ESD interfere with the operation
of equipment, counter measurements such as
wrist strap, grounding shall be considered.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
N/A
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
N/A
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
(100 % dip in
UT) for 0,5 cycle
0 % U
T
(100 % dip in
UT) for 1 cycles
70 % U
T
(30 % dip in
UT) for 25/30cycles
0 % U
T
(100 % dip in
N/A
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the FDTH3400 requires continued
operation during power mains interruptions, it
is recommended that the FDTH3400 be
powered from an uninterruptible power
supply or a battery.