measurements 

 

Note: Comparative blood pressure measurements should be performed in the same condition 

(usually referred to quiet conditions)! 

 

  If the arm artery is low (high) relative to the heart , it will result in a high (low) blood pressure value. (Each 15cm height 

difference will produce 1.3kPa/mmHg error value)! 

 

  This unit is not suitable for newborns. 

 

  Note: Please use the original arm cuff that meets the clinical test requirement! 

 

  The loose or air bag exposed cuff will lead to false blood pressure. 

 

  The arm will compress the blood vessels due to repeated measurements. This condition will also cause a biased blood 

 

pressure value. Therefore, when making repeated measurements, be sure to rest for 3-5 minutes or raise your arm for 3 minutes 

to reduce the congestion.

 

 

8.Measuring blood pressure                                                                       

 

8.1 Start measuring   

Press the [start/stop] key to start automatic measurement. 

 

 

 

 

 

 

 

 

 

 

 

Содержание FDBP-A10BL

Страница 1: ...anks for using our product Take care of this instruction manual Please read this manual carefully before using Precaution Please read the instruction manual carefully before use it If you have any questions please consult the salesperson in the purchasing center or ask your doctor FCC ID PONFDBP ...

Страница 2: ... pressure 4 9 View measurement results 4 10 Set the function 5 11 WHO blood pressure definition function 6 12 Maintenance and storage 6 13 Blood pressure measurement Q A 7 14 Blood pressure knowledge 7 15 Error warning Troubleshooting 8 16 Reference standard 9 17 Place in 9 18 Technical specifications 9 19 List of attachment 10 20 Standardized symbolic description 10 21 Electromagnetic compatibili...

Страница 3: ...1 Quick instructions 2 Product Structure Description 2 1 Upper Arm Blood Pressure Monitor 2 2display ...

Страница 4: ...g is inflated for a long time immediately press the START STOP button to stop the measurement Continued inflation presses the hand arm which can cause limb numbness Do not allow children to use the instrument without authorization Some small parts may cause ingestion It is forbidden to measure blood pressure on the side where the breast is removed Do not allow children to use instruments without p...

Страница 5: ...d pulse rate of adults and children at least 12 years of age at household or medical center Not suitable for neonate pregnancy or pre eclampsia with the cuff around the left upper arm according to the instruction in the user s guide manual the patient is an intended operator Intended operator should receive at least 8 years of education so that can read and understand the user manual 4 Instruction...

Страница 6: ...ote Consult the dealer at the place of purchase for any adapter problems 5 3Use adapter Except the battery this unit can also use a adapter USB 5V Must use the original or medical power supplies that meet medical standards example IEC 60601 1 certification Transformer connects to power socket is shown on the following Adaptor connects to the power socket of the product and press the START STOP mem...

Страница 7: ... sit in the chair place your feet flat on the floor place your arms on the table and put the arms and the heart on the same level to relax and measure in a relaxed and natural posture 7 Correct measurement posture Preparation before measurement For wireless features see the attachment A v o i d e a t i n g s m o k i n g o r any form of activity before measuring all of which will affect the measure...

Страница 8: ...le for newborns Note Please use the original arm cuff that meets the clinical test requirement The loose or air bag exposed cuff will lead to false blood pressure The arm will compress the blood vessels due to repeated measurements This condition will also cause a biased blood pressure value Therefore when making repeated measurements be sure to rest for 3 5 minutes or raise your arm for 3 minutes...

Страница 9: ...on The display represents speaking or moving during the measurement The measurement may not be correct Please measure again 3 Cuff wearing self test prompt icon When the cuff is wrapped correctly the screen displays When the cuff is wrapped too loosely the screen displays Please press the START STOP button to turn off the power rewind the cuff correctly and measure again 8 3 Removing the Arm Strap...

Страница 10: ...e Monitor can only delete all records at one time Cannot delete a single record When the measurement result is displayed press and hold the MEM key while pressing the start stop key for about 3 seconds Please do not press the start stop key first otherwise the power will be cut off 10 Set the function The system defaults to user ...

Страница 11: ...ing mode if not selected the system defaults to the manufacturing time The last two digits of the year will flash on the screen You can enter the year by pressing the MEM button for example you can increase the year from 21 to 22 by pressing it one time 2 Press the SET button and there will be a jump of the month in the left corner of the screen At this time the first digit of the month starts to ...

Страница 12: ... time by pressing the MEM button for example you can change the time from 0 00 to 12 00 by pressing the MEM button 12 times 5 Press the SET button again and the last two digits in minutes will start to flash You can enter the minute by pressing the MEM button for example pressing the button 8 times will increase the time from 0 minutes to 8 minutes ...

Страница 13: ...shut down after 30 seconds 11 WHO blood pressure definition function This unit comes with a SYS warning and a warning bar reminder function we can read from the height of the World Health Organization blood pressure indicators blood pressure is distributed in DIA no WHO voice prompts and ideal blood pressure normal blood pressure green normal partial hypertension yellow mild hypertension moderate ...

Страница 14: ...ect measurement Clean When the product is dirty due to prolonged use it is necessary to follow the requirements for cleaning When the outer shell is dirty gently wipe it with a wet soft cotton cloth If the outer shell is seriously soiled wipe it with a soft dry cloth dipped in 75 medical alcohol It is recommended to clean at least once a month Please pay attention to keeping the cuffs clean After ...

Страница 15: ...emble or adjust this unit if you feel the product has functional problems Electronic Upper Arm Blood Pressure Monitor is a very precise product any improper maintenance disassembly and adjustment will lead to inaccurate measurement Avoid the damage of cuff and rubber tube caused by folding Avoid falling or shaking this unit violently Within warranty period please contact the distributor or manufac...

Страница 16: ...ion the blood pressure monitor has a hard time picking up arterial signals and the blood pressure readings are too high Please confirm again that the binding position of the armband is correct A Wrapping the armband too loosely can sometimes lead to high measurements If the band is wrapped too loosely the pressure does not reach the arteries so the blood pressure is measured higher than it actuall...

Страница 17: ...Q When is the best time to measure blood pressure Answer The best time to measure blood pressure is within 1 hour after getting up in the morning or before going to bed at night It is also recommended to measure at the same time every day 14 Blood pressure knowledge 14 1High and low pressure The highest blood pressure when the heart contracts and sends blood to the arteries is called high pressure...

Страница 18: ... including 1 Dietary Habit Maintain normal weight as guided by doctor Do not eat too much salt as many packaged foods contain more salt Avoid eating greasy food Packaged foods usually contain large amounts of fat 2 Preventing Diseases Adhere to medical guidelines for preventing certain diseases such as diabetes fat metabolic disorders and gout 3 Living Habit Don t smoke avoid drinking too much or ...

Страница 19: ...or 18s did not reach 25mmHg 5 After the air pressure exceeds 30mmHg the air pressure is lower than 20mmHg due to the looseness of the cuff Adjust the cuff It is advisable to insert two fingers just after the tightness is tied The cuff mouth of the cuff is kept 2cm on the elbow socket Lo Power is not enough The power is lower than the minimum operating voltage Replace the battery Over range alarm W...

Страница 20: ...ic ring falling off Please put the plastic ring on and measure again Any technical problems related to blood pressure monitor should consult experts or medical personnel do not disassemble repair without permission Unauthorized disassembly will lose the warranty right 16Reference standard 参考标准 We Famidoc Technology Co Ltd solemnly declares that the FDBP A10BL Upper Arm Blood Pressure Monitor we pr...

Страница 21: ...20mm Software version V1 0 Electric shock protection type Class II device Shock protection procedure BF application part Expected use lifetime Body in 5 years Cuff in 2 years Application component Cuff Operating condition Temperature 5 C 40 C Humidity 15 RH 90 RH No condensation Atmospheric pressure 70kPa 106kPa Transportation and storage condition Temperature 25 C 55 C Humidity 15 RH 95 RH No con...

Страница 22: ...20 Standardized symbolic description ...

Страница 23: ...to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This transmitter must not be co located or operated...

Страница 24: ...10BL is intended for use in the electromagnetic environment specified below The customer or the user of FDBP A10BL should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The FDBP A10BL uses RF energy only for its internal function There for its RF emissions are very low and are not likely to cause any interf...

Страница 25: ...ronment If the user of the FDBP A10BL requires continued operation during power mains interruptions it is recommended that the FDBP A10BL be powered from an uninterruptible power supply or a battery Power frequency 50 60Hz magnetic field IEC 61000 4 8 30 A m 30 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environme...

Страница 26: ...ng to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey b should be less than the compliance level in each frequency range c Interference may occur in the vicinity of equipment marked with the following symbol Proximity magnetic fields IEC 61000 4 39 30 kHz CW 8A m This te...

Страница 27: ...le and mobile RF communications equipment and the EQUIPMENT or SYSTEM for EQUIPMENT and SYSTEMS Recommended separation distances between portable and mobile RF communications equipment and the FDBP A10BL The FDBP A10BL is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the FDBP A10BL can help prevent electromagnetic in...

Страница 28: ...al Zone Jinxia Changan Town Dongguan 523853 Guangdong Province P R China Tel 86 769 89272488 Website www famidoc com Name Shanghai International Holding Corp GmbH Europe Add Eiffestrasse 80 20537 Hamburg Germany PN Description Version V1 0 Revision date April 2022 ...

Отзывы: