measurements
Note: Comparative blood pressure measurements should be performed in the same condition
(usually referred to quiet conditions)!
If the arm artery is low (high) relative to the heart , it will result in a high (low) blood pressure value. (Each 15cm height
difference will produce 1.3kPa/mmHg error value)!
This unit is not suitable for newborns.
Note: Please use the original arm cuff that meets the clinical test requirement!
The loose or air bag exposed cuff will lead to false blood pressure.
The arm will compress the blood vessels due to repeated measurements. This condition will also cause a biased blood
pressure value. Therefore, when making repeated measurements, be sure to rest for 3-5 minutes or raise your arm for 3 minutes
to reduce the congestion.
8.Measuring blood pressure
8.1 Start measuring
Press the [start/stop] key to start automatic measurement.