Thermographic measurement techniques
3. Measure the radiation intensity (= apparent temperature) from the reflecting source us-
ing the following settings:
• Emissivity: 1.0
• D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
Figure 35.3
1 = Reflection source
Figure 35.4
1 = Reflection source
Using a thermocouple to measure reflected apparent temperature is not recommended for
two important reasons:
• A thermocouple does not measure radiation intensity
• A thermocouple requires a very good thermal contact to the surface, usually by gluing
and covering the sensor by a thermal isolator.
35.2.1.1.2
Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
#T559954; r. AL/37426/37426; en-US
231
Содержание T10 series
Страница 1: ...User s manual FLIR T10xx series ...
Страница 2: ......
Страница 3: ......
Страница 5: ...User s manual FLIR T10xx series T559954 r AL 37426 37426 en US v ...
Страница 6: ......
Страница 14: ......
Страница 209: ...Technical data 26 72501 0402 FLIR T1050sc 28 72501 0403 FLIR T1050sc 45 T559954 r AL 37426 37426 en US 195 ...
Страница 217: ...Mechanical drawings 27 T559954 r AL 37426 37426 en US 203 ...
Страница 226: ...Digital I O connection diagram 29 T559954 r AL 37426 37426 en US 212 ...
Страница 227: ......
Страница 228: ...CE Declaration of conformity 30 T559954 r AL 37426 37426 en US 214 ...
Страница 229: ......
Страница 236: ...Application examples 32 T559954 r AL 37426 37426 en US 222 ...
Страница 275: ......