FTP3000 Series Programmable DC Power Supply Operation Manual
____________________________________________________________________________
- 14 -
2.6 Installing
2.6.1 Preparation For Use
(1) Ensure the Power Supply is connected to the AC line input that meets the specification.
(2) The instrument must be installed in an area with good air circulation to avoid the internal temperature getting
too high.
(3) For indoor use only, ambient temperature 0°C to 40°C.
2.6.2 Requirements of Input Power
Please pay attention to the following items for FTP3000 series power supply input connection:
①
Single phase AC input: 190VAC~265VAC, 47~63Hz;
②
The AC input power cord is a 3-core cable with grounding connection, DO NOT operate the instrument if there
is no proper socket;
③
There is 1PCS AC input power cable included with the accessories, which should be compatible with your local
electric voltage. Contact Faithtech agent or after-sale service immediately if you find the input cable is not
matching with the power supply AC input terminal.
----------------------------------------------------------------------------------------------------------------------------
The instrument provides mainframe grounding connection via 3-core power cable. Make
sure there is proper ground connection in applying socket before operation. Under no
circumstances shall this DC Power Supply be operated without an adequate ground
connection.
----------------------------------------------------------------------------------------------------------------------------
2.6.3 Power ON Self Check
When powered on, the power supply will conduct self check, the check process and results will be displayed on
the screen, such as below:
Figure 2-5 FTP3000 Power ON Self Check
The power supply will enter MAIN PAGE if all checking results are OK, otherwise it will wait for 3 seconds before
entering MAIN PAGE.
If there is FAULT during power supply power on self check, please refer to “Table 5-2 Power On Self-Check Fault”.