9
To enter the configuration section, from the
sub-menu, press this button.
In the configuration menu the top row has 3
buttons available, from left to right these are:
Set Date
Set Time (the instrument uses a 24 hours
clock)
Set units of measure and date format.
The date and time are set prior to the unit
being supplied but they can be adjusted as
necessary.
5.3 USING THE TEST PORT
The F6000 test port can also be used a range with the following Impactors/tubes listed below:
Test
Draeger ref. Factair Part
No:
Sensitivity
Oil (Impactor)
8103530
TUB0053
0.1 mg/m
3
Oil (chemical reagent
tube)
6728371
TUB0003
0.1 mg/m
3
Nitrous Fumes
NO
x
8103661
TUB0068
0.2-6 ppm
Sulphur Dioxide
SO
2
6728491
TUB0014
1-25 ppm
Polytest
CH28401
TUB0017
Sensitivity
dependant
on
compound present
To prepare the tubes Factair recommends the F2187 Draeger tube tip cutter, this is not
supplied as standard.
If you have another Draeger tube, which you wish to use with the instrument, contact Factair
for technical advice on its suitability.
To complete a tube or Impactor test from the home screen click on the arrow in the top right
hand corner then press the configuration menu button.
Draeger Oil Impactor Test