Factair F6100 Скачать руководство пользователя страница 1

 

 

 

 

F6100 SAFE-AIR MONITOR 

OPERATING INSTRUCTIONS 

 

 

 

 
FAC QAM 097 Issue E 03/21 

 

 

Содержание F6100

Страница 1: ...0 F6100 SAFE AIR MONITOR OPERATING INSTRUCTIONS FAC QAM 097 Issue E 03 21...

Страница 2: ...1...

Страница 3: ...ART UP 7 5 2 F6100 SCREEN MENUS 8 5 3 USING THE TEST PORT 9 6 F6100 PC SOFTWARE 13 6 1 INSTALLING THE SOFTWARE 13 6 2 USING THE SOFTWARE 13 7 DATALOGGED RESULTS 17 APPENDICES Appendix 1 Helpful tips 1...

Страница 4: ...providing the concentration in mg m3 It also includes digital airline pressure and ambient temperature readings Periodic oil measurements are completed using the Draeger Impactor which is inserted int...

Страница 5: ...61 TUB0068 0 2 6 ppm Sulphur Dioxide SO2 6728491 TUB0014 1 25 ppm Polytest CH28401 TUB0017 Sensitivity dependant on compound present The tester can also be used with the old type Draeger reagent tube...

Страница 6: ...D Training Video d USB Memory Stick with PC Power On Air Inlet Volt free Contact for remote alarm The current ratings are 0 5A at 125Vac and 1A at 24Vdc Mains input USB Connection for PC Software Wall...

Страница 7: ...tions on the wall and check with a spirit level Ensure you use the correct fixings to suit the wall Once you have installed the bracket mount the F6100 in place and connect the power supply It is impo...

Страница 8: ...s calibrated the instrument will then display the home screen At this point you can then connect the breathing air supply to the air inlet Please note that the tester can be piped to any pressure betw...

Страница 9: ...for the configuration settings or to perform a Draeger Impactor or tube test press the arrow key in the top right hand corner The sub menu displays the date when the test port was last used and the r...

Страница 10: ...Factair Part No Sensitivity Oil Impactor 8103530 TUB0053 0 1 mg m3 Oil chemical reagent tube 6728371 TUB0003 0 1 mg m3 Nitrous Fumes NOx 8103661 TUB0068 0 2 6 ppm Sulphur Dioxide SO2 6728491 TUB0014...

Страница 11: ...sted Once the purge has finished the Draeger Impactor should be inserted into the test port and tightened firmly ensure the Impactor has an intact protective seal This seal must remain in place for th...

Страница 12: ...from oil sulphur dioxide nitrous fumes or the polytest tube If you are intending to complete an oil tube test then you will need to input the test time If you can determine with a high degree of cert...

Страница 13: ...o the test Ensure the arrow printed on the tube is pointing away from the instrument Press the arrow in the top right corner to proceed The test will proceed as described in section 3 and at the end o...

Страница 14: ...alled when you run the application the following screen will be displayed Once installed on your computer run this program and then with the USB cable provided connect your PC or laptop to the instrum...

Страница 15: ...yed to confirm you have connected the cable press ok to proceed A message will then be displayed to confirm the settings have been downloaded into the software click ok to proceed A typical example of...

Страница 16: ...nternal Sensors In the Internal Sensors section you can select from a range of international breathing air standards The menu display the current test standard and you will be able to select a range o...

Страница 17: ...ther assistance in configuring these sockets please contact Factair Updating the F6100 settings Once you have made all the required changes click on the Write to F6100 button A message will then be di...

Страница 18: ...e website F6100 Data Monitoring Software http www factair co uk downloads The datalogging software makes it easy to analyse results identify trends and exceptions Results can be imported into the soft...

Страница 19: ...strument off press the button in the top right hand corner of the screen The following warning will be displayed if there is insufficient pressure being supplied for the tube test The instrument requi...

Страница 20: ...ally designed to prevent glass from falling out of the opener by accident The reservoir for the broken off tips is easy to empty Instructions 1 Place the end of the tube between the 3 blades and turn...

Страница 21: ...and quickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remain blank For tests above that read as follows The bottom line re...

Страница 22: ...ll show a light brown or darker discoloration Synthetic oil The white crystals will show a yellow discoloration Note best seen by comparing with an unused tube 1 Using the Tube Tip Cutter place the am...

Страница 23: ...ting the tube reading is multiplied by a factor of 10 i e a tube reading of 150mg m becomes 1500mg m To establish the pressure dewpoint refer to the graph on the back of the Safe Air Tester Result She...

Страница 24: ...oxide The total length of the white section is the measure of concentration read directly from the 1 25ppm scale in parts per million Polytest CH 28401 REQUIREMENT NO DISCOLORATION The media in the de...

Страница 25: ...ill or incineration OR 2 Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles Dispose of the bin via the company s normal hazardous waste disposal method s ie landf...

Страница 26: ...25...

Страница 27: ...26 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Fax 44 0 1473 747123 Email enquiries factair co uk www factair co uk...

Отзывы: