COSHH L5 (Sixth Edition 2013)

 

EXTRACTS FROM CODE OF PRACTICE RELATING TO 

RESPIRATORY PROTECTIVE EQUIPMENT (RPE) 

178. The maintenance, examination and  tests should be in accordance with the manufacturer's instructions. 
Examinations  should  comprise  a  thorough  visual  examination  of  all  parts  of  the  respirator  or  breathing 
apparatus, to ensure that all parts are present, correctly fltted, and the equipment is in good working order. In 
particular, the examination should ensure that the straps, facepieces, filters and 

valves are sound and in good working condition. For powered and power-assisted respirators, tests should: 

• be made on the condition and efficiency of those parts; 

• ensure that the battery pack is in good condition; 

• ensure that the respirator delivers at least the manufacturer's recommended minimum volume flow rate. 

179. For RPE incorporating compressed gas cylinders, tests should include the condition and efficiency of all 
parts, the pressure in the cylinders and the volume flow rate. Frequency of examination and tests 

180. The quality of the air supplied to a breathing apparatus should be tested at suitable intervals, depending 
on the task and the frequency of use. When the air supply is from mobile compressors, the employer should 
ensure  that  wherever  a  compressor  is  located,  the  quality  of  air  it  supplies  is  not  compromised  by  nearby 
contaminants.  In  every  case,  the  air  supplied  to  a  breathing  apparatus  should  meet  the  relevant  quality 
standard.  As  it  is  not  reasonably  practicable  to  test  for  all  contaminants,  the  risk  assessment  made  under 
regulation 6 should guide what other contaminants will require testing. 

 

181. Thorough maintenance examinations and, where appropriate, tests of items of RPE, other than one-shift 
disposable respirators, should be made at suitable intervals. The frequency should increase where the health 
risks and conditions of exposure are particularly severe. 

 

182.  In  situations  where  respirators  are  used  only  occasionally,  an  examination  and  test  should  be  made 
before  their  next  use  and  maintenance  carried  out  as  appropriate.  The  person  who  is  responsible  for 
managing  the  maintenance  of  RPE  should  determine  suitable  intervals  between  examinations.  Emergency 
escape-type RPE should be examined and tested in accordance with the manufacturer's instructions. 

 

183. Suitable arrangements should be made to ensure that no employee uses RPE which has previously been 
used  by  another  person,  unless  it  has  been  thoroughly  washed  and  cleaned  in  accordance  with  the 
manufacturer's instructions. 

 

 

Содержание F4504

Страница 1: ...0 F4504 MULTI AIR TESTER OPERATING MANUAL FAC QAM 173 Issue G 03 21...

Страница 2: ...1...

Страница 3: ...iewing Test Results 24 9 Exporting Test Results and PC Software 25 Appendix 1 Helpful Tips 27 Appendix 2 Draeger Tube Tip Cutter 28 Appendix 3 Reading the Draeger Oil Impactor and Chemical Reagent Tub...

Страница 4: ...European Phamacopeia Medical and Surgical Air European Phamacopeia CO2 European Phamacopeia N2 European Phamacopeia N2O The F4504 has a colour touch screen display and is supplied in a hard weatherpr...

Страница 5: ...reagent tubes Impactor Test Draeger ref Factair Part No Sensitivity Carbon Monoxide CO 6728511 TUB0004 5 150 ppm Carbon Dioxide CO2 6728521 TUB0005 100 3000 ppm Oil Impactor 8103560 TUB0053 0 1 mg m3...

Страница 6: ...1946 Connection BSP inlet adaptor Stylus Pen Optional Extras F3005 20 Bar Regulator F3002 200 300 Bar 5 8 DIN thread High Pressure Regulator F2193 Connection adaptor terminating in a CEJN Plug Sintere...

Страница 7: ...mains power supply To install the alkaline AA batteries lift the tester out of its case open the battery drawer at the rear of the unit and place the batteries inside Note It is important to strictly...

Страница 8: ...on period The F4504 series has 2 battery level displays the left hand one is for the alkaline AA batteries if fitted The right hand one is the rechargeable internal lithium battery Unless the instrume...

Страница 9: ...ther details 3 Selecting the test standard and configuring the instrument From the main menu press the configure icon In the configuration sub menu there are 3 icons available from left to right these...

Страница 10: ...er to exit To then exit from the configuration sub menu to the main menu press the arrow at the top left of the screen 4 Breathing Air Tests All Having selected the breathing air standard EN12021 in t...

Страница 11: ...ternative chemical reagent tubes see appendices 2 and 3 For breathing air tests the standard option is to select a Draeger oil Impactor The unit has 4 breathing air test programmes available Airline S...

Страница 12: ...eturned to the manufacturer for checking and resetting Input pressure will be displayed on the measured pressure bar graph and the adjacent display System pressure will automatically default to this r...

Страница 13: ...there is a serious odour problem fail the sample at this point of the test by pressing the fail button Initiating flows of seriously contaminated air could result in damage to the instrument If the od...

Страница 14: ...ment detects that the dewpoint meter s rate of change is still too high 0 2 dec C per minute it will automatically continue the test until this stabilises maximum test time 15 minutes Once the test is...

Страница 15: ...d test where carbon dioxide failed to meet the requirements of the selected standard For airline systems less than 40 bar the pressure dewpoint i e the point at which water will begin to condense into...

Страница 16: ...d fully charged cylinder Note For the high pressure charging compressor test a DIN to DIN adaptor ref F2158 is required to connect the F3002 high pressure regulator to a charging hose To carry out a t...

Страница 17: ...ssed through the test port and across the electronic cells This clears contaminants from previous tests and ensures that a representative sample is being tested Once the purge has finished the Draeger...

Страница 18: ...f the manual The test complete screen displays all the measurements recorded during the test And automatically determines whether the test was a pass or fail against the standard displayed in the top...

Страница 19: ...r than 40 C inside the instrument the calibration will operate for an extended period to stabilise the Oxygen cell You now have the option of entering a description for the test location if you wish t...

Страница 20: ...will automatically default to this reading Note The pressure dewpoint at the end of the test will be calculated on the recorded system pressure For the majority of applications the system pressure ca...

Страница 21: ...gether with test standard The pressure ambient temperature and oxygen content will also be displayed During the test the flowmeter should be kept closed Test port volumes are automatically adjusted fo...

Страница 22: ...f the screen For the oil and polytest tubes select the appropriate pass or fail icons After you have entered each result that box will change from red to green Once you have completed entering all the...

Страница 23: ...t and only disconnect at the end of the supply hose The adaptor can then be disconnected afterwards if required To turn off the Tester use the return arrows to return to the main menu screen and press...

Страница 24: ...calibration mode Note If the temperature is greater than 40 C inside the instrument the calibration will operate for an extended period to stabilise the Oxygen cell The instrument will then change op...

Страница 25: ...and time when the test was completed To review a previous test select it then press the right hand arrow To delete a test highlight it then press the trash can button All tests can be deleted by selec...

Страница 26: ...in the download section of Factair s website http www factair co uk downloads Then click on the File drop down menu and select import test results from F4500 via USB connection You will then be able t...

Страница 27: ...the RPD by selecting pass or fail as appropriate Note In addition there is an option of entering N A against these categories In this instance the resultant test certificate will only validate air qua...

Страница 28: ...tube tip cutter ensure only a light pressure is applied to the tube whilst rotating for scoring General Make sure the system flow control valve is closed after each test Connecting an air supply to th...

Страница 29: ...s from falling out of the opener by accident The reservoir for the broken off tips is easy to empty Instructions 1 Place the end of the tube between the 3 blades and turn to score the end 2 Push the t...

Страница 30: ...y vents around its circumference this allows the user to easily and quickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remai...

Страница 31: ...ed test Mineral oil The white crystals will show a light brown or darker discoloration Synthetic oil The white crystals will show a yellow discoloration Note best seen by comparing with an unused tube...

Страница 32: ...om the printed scale at that point is a measure of the concentration in parts per million For the CO2 tube 67 28521 there is an option to adjust the test volume up by a factor of two to increase the n...

Страница 33: ...tube 67 28511 has a range of 0 to 150 PPM and the maximum limit for medical air in Pharmacopeia is 5 PPM An alternative tube 67 33051 shown below which has a 0 to 60PPM range can also be used with the...

Страница 34: ...section is the measure of concentration read directly from the 1 25ppm scale in parts per million Polytest CH 28401 REQUIREMENT NO DISCOLORATION The media in the detector tube will discolour to provid...

Страница 35: ...ccordance with local authority waste regulations some tube aqueous waste may require neutralisation prior to disposal Place the tubes in a sharps or glass bin wearing protective gloves and safety spec...

Страница 36: ...the apparatus is used and stored at a known temperature the pressure dewpoint shall be at least 5 C below likely lowest temperature Where the conditions of usage and storage of any compressed air supp...

Страница 37: ...cooling will result in air entering the filtration at too high a temperature this will cause the filtration to prematurely fail and pass excess levels of oil and water Malfunctioning dryers can distur...

Страница 38: ...ty is tested whenever it is first moved into a new position or prevailing wind conditions change The final decision on frequency of test is the responsibility of the Employer and needs to not only ref...

Страница 39: ...e condition of the equipment and details of any defect found including for canister or filter respirators the state of the canister and the condition of the filter for self contained compressed air ga...

Страница 40: ...quality of the air supplied Table A 2 provides a summary of the main elements associated with a compressor system for producing breathable air In addition to the careful and installation of the system...

Страница 41: ...quency of use When the air supply is from mobile compressors the employer should ensure that wherever a compressor is located the quality of air it supplies is not compromised by nearby contaminants I...

Страница 42: ...is not in use The adequacy of the accommodation will vary according to the quantity type and its use eg pegs labelled lockers shelves or containers etc The storage should be adequate to protect the PP...

Страница 43: ...compatible with the BA Never place the hose inlet near to potential sources of contamination eg vehicle exhausts Never use the equipment without the waist belt Never use a light duty airline hose whe...

Страница 44: ...to as low a level as possible and shall be far below the national exposure limit Combination effects of more than one contaminant shall be taken into account 1999 Respiratory protective equipment at...

Страница 45: ...scoloration No discoloration Odour None None Note There is also a requirement to conduct a visual particulate test and this should be free from visible particles in a 75 L sample 175 L for surgical ai...

Страница 46: ...45 Water mg m3 to PPM v v Conversion Charts...

Страница 47: ...46...

Страница 48: ...47 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Fax 44 0 1473 747123 Email enquiries factair co uk www factair co uk...