Service schedule
Service Number
1
2
3
4
5
6
7
8
Tires:
Slack, wear, check the tires
Gas Springs:
Check for wear
Mounting Mechanism:
Slack, wear
Throttle:
Operation, battery, speed setting
Electric Mechanism:
Wing, affiliation
Brushes of the engine
Check by (initials)
Date
7.
Problem analysis and solutions
It could be there are problems when using your Click & Go Professional II. The potential
problems that may occur are described below in the table. You will also find the solutions.
These solutions must be performed by an authorized Faaborg Rehab dealer. Only at code 1
you can do something yourself. If you have problems with your push propulsion that cannot
be found in the table below, we recommend you to contact an authorized Faaborg Rehab
dealer.
Code
Problems
Solutions
1
Batteries to empty
Charge the battery.
2
Bad contact motor
Make sure that the connector of the motor is connected correctly and
that the brushes make contact.
3
Short circuit motor
Check the motor and replace the controller.
4
Not applicable
-
5
Not applicable
-
6
No reaction
The charger is connected. Also check the inputs of the push propulsion.
7
Potentiometer error
Check the potentiometer and the wiring.
8
Controller error
Check the connectors and wiring. If necessary, replace the controller.
9
Parking brake error
The push propulsion is still on the free wheel or the wiring makes
poor contact.
10
Bad nutrition
Make sure the batteries are properly connected and the wiring makes
good contact.