F Care Systems EVRF Скачать руководство пользователя страница 2

       D-COM-QUGe-07

EVRF settings for the CR45i catheter

Diameter vein

6 - 18  mm

Catheter

CR45i

Needle

6 French Introducer

Power (watt)

25

Continuous mode

Keep footswitch pressed down

Number beeps at start

2 x 3

Number beeps per 0,5 cm  

retraction

3

Connection of pad required

YES

Parameters for treatment of varicose veins and hemorrhoids:

www.fcaresystems.com - Tel: +32 3 451 51 45     

EVRF settings for the transcutaneous treatment

Diameter vein

0,3 - 0,6 mm

0,7 - 1 mm

Treatment area

Face

Leg

Needle

K3i

K6i

Power (watt)

5 - 8 

Max. 10

9 - 14  

Max. 16

Impulse time

0,2 - 0,3 s

0,3 - 0,4 s 

Max 0,5s

Connection of pad required

NO

NO

1639

EVRF settings for the HPR45i probe

Hemorrhoids

Degree II, III, IV

Probe

HPR45i

Power (watt)

25

Continuous mode

Keep footswitch pressed down

Connection of pad required

NO

EVRF settings for the CR30KAB catheter

Diameter vein

2 - 5mm

Catheter

CR30KAB

Needle

G20 needle

Power (watt)

15

Continuous mode

Keep footswitch pressed down

Number beeps at start

3

Number beeps per 0,8cm  

retraction

3

Connection of pad required

Optional