84
Operation
Operation
Manual
Manual
2. Pre s s (R) o r (L) to c ho o s e the fie ld fo r te s t.
3. On the touch screen, press [CC 0.25] [CC 0.50] button to change it into Jackson
Cross Cylinder Mode.
4. While pressing number
⑴
and number
⑵
, ask the patient
which side letter is seen more clearly.
▶
If number 1 is seen clearly, turn the dial counterclo direction).
▶
If number 2 is seen clearly, turn the dial clockwise(- direction).
5. To estimate the distorted vision axis, press [A] or left/right button. To estimate
circumference power, press [C] or left/right button.
[Figure 10-12] Jackson Cross Cylinder Test (Distorted Vision Axis Test)
10.13
Dua l Cro s s Cylin d e r Te s t
!
Purpose: In subjective test, revise maximum circumference refractive power and axis for
Monocular, using Dual Cross Cylinder
!
Chart: Dot-Group Chart
!
Auxiliary Lens: Dual Cross Cylinder Lens (DUAL)
!
Target: Separted two Dot-Group Chart must be seen samely.
!
Way of Test
1. On Special Chart clause of touch screen, press Dot-Group Chart to move on to
Cross Cylinder Mode.
2. Press (R) or (L) to choose the field for test.
Содержание EDR 9000
Страница 1: ... EDR 9000 Digital Retractor Operation Manual ED R 9000 Digital Retractor ...
Страница 2: ......
Страница 11: ...9 Appendix Hardware Specification 115 Digital Refractor EDR 9000 ...
Страница 64: ...62 Operation Operation Manual Manual Figure 9 2 Adjustment of 1 st step S C A Value ...
Страница 71: ...69 Figure 9 14 Far Distance Vertiacl Coincidence Aniseikonia Test Digital Refractor EDR 9000 ...
Страница 118: ...116 Operation Operation Manual Manual ...