10. FAULT FINDING
Hereunder are some types of disorders, which can usually be repaired rather simp-
ly. If these following measures are unsuccessful, contact your authorized dealer!
Problem
Check Point
The driving of the wheelchair is
too slow or insufficient, the status
indicator lamp flashes slowly
Batteries are exhausted and need to be
recharged
The
Ezee Travel
does not function,
the Battery/Status
indicator
flashes
rapidly.
Count the number
of flashes:
1 flash
The battery needs charging or there is a
bad connection to the battery. Check the
connections to the battery or charge bat-
teries.
2 flashes
There is a bad connection to the motor.
Check all connections between the mo-
tor and the controller.
3 flashes
The motor has a short circuit to a battery
connection.
4 flashes
Freewheel is engaged.
5 flashes
Not used.
6 flashes
The controller is being inhibited from
driving; this may be because the battery
charger is connected.
7 flashes
A throttle fault is indicated. Make sure
that the throttle is in the rest position be-
fore switching on the scooter
8 flashes
A controller fault is indicated. Make sure
the controller connections are secure.
9 flashes
The parking brakes have a bad
connection. Check the parking brake
and motor connections. Make sure the
controller connections are secure.
10 flashes
An excessive voltage has been applied
to the controller. This is usually caused
by a poor battery connection. Check the
battery
packs
and
controller
connections.