CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 33 of 49
Figure 40: Rough positioning of the device to the patient
The fine positioning will performed during initiation of treatment. Please follow section 4.7.2.
4.7
Treatment
Perform an illumination check (Procedure see section 4.5) prior to each treatment.
For this purpose use the attached measurement adapter and the included UV
–
meter. The check makes sure that the irradiation intensity is in its defined limits and
no harm is caused to the patient.
4.7.1
Start of treatment
Have a clock ready at the start of the treatment and control the timing of the treatment. The
treatment is started by pressing the Start / Stop button. The device displays during treatment in the
display "EMISSION", and the emission light is on. The remaining time of the treatment is in the 2nd
display line (Figure 41).
Figure 41: Display while treatment
Check proper functioning of the device and proper positioning above the patients eye
(correct distance).
For errors that occur on the device or unclear device states turn the device off and
contact your distributor for repair or replacement of the device
Distance between bottom of the
unit and the eye = 50 ± 5 mm