background image

DO600-en-EU_V7.2   09/14 

11

Appendices

 

Troubleshooting

 

Guide

 

Symptom

 

Possible

 

Causes

 

Action

 

Unit

 

will

 

not

 

power

 

on

 

Batteries

 

not

 

in

 

place

 

Dead

 

batteries

 

Battery

 

polarity

 

incorrect

 

Replace

 

Batteries

 

Replace

 

Batteries

 

Reorient/Replace

 

Batteries

 

“BAT”

 

indicator

 

shown

 

on

 

display

 

Batteries

 

are

 

weak

Replace

 

Batteries

 

Unstable

 

Readings

Insufficient

 

electrolyte

 

in

 

probe

 

(Air

 

bubbles

 

present

 

when

 

probe

 

is

 

inverted.)

 

Electrolyte

 

is

 

depleted

 

Replace

 

electrolyte,

 

and

 

membrane

 

cap

 

assembly.

 

Readings

 

drift

 

down

Insufficient

 

stirring

 

(the

 

probe

 

consumes

 

oxygen

 

at

 

the

 

measuring

 

surface,

 

requiring

  

constant

 

sample

 

movement

 

across

 

membrane)

 

Move

 

probe

 

in

 

sample

 

or

 

stir

 

sample

 

Slow

 

Response

Dirty

 

or

 

damaged

 

membrane

 

Replace

 

electrolyte

 

and

 

membrane

 

cap

 

Electrode

 

cannot

 

be

 

calibrated

 

Depleted

 

electrolyte

 

Dirty

 

or

 

damaged

 

membrane

 

Replace

 

electrolyte

 

and

 

membrane

 

cap

 

Electrode

 

cannot

 

be

 

calibrated

 

after

 

replacing

 

the

 

electrolyte

 

and

 

membrane

 

cap

 

Dirty

 

probe

 

(Cathode

 

is

 

not

 

a

 

shiny

 

gold

 

color)

 

Clean

 

cathode

 

with

 

Polishing

 

paper

 

*

 

perform

 

the

 

electroninc

 

Zero

 

calibration

 

Sample

 

reading

 

is

 

frozen

 

Unit

 

is

 

in

 

“HOLD”

 

mode

 

Unit

 

is

 

locked

 

Release

 

HOLD

 

(momentarilly

 

press

 

Mode/Hold

 

button)

 

Remove

 

batteries,

 

press

 

ON/OFF

 

button,

 

replace

 

batteries

 

and

 

restart

 

* Polishing

 

paper

 

is

 

available

 

in

 

the

 

DO603

 

membrane

 

kit

 

netzerotools.com

netzerotools.com

Содержание ExStik DO600

Страница 1: ...Dissolved Oxygen DO600 II RECALL USER GUIDE ExStik DO600 Dissolved Oxygen Meter netzerotools com netzerotools com netzerotools com Extech DO600 K Waterproof ExStik II Dissolved Oxygen Meter Kit ...

Страница 2: ...e compensation This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Please visit our website to check for the latest version of this User Guide Product Updates and Customer Support Meter Description Front Panel Description 1 Battery compartment cover 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Retaining Co...

Страница 3: ...lapse before measurements can be made A special circuit is included in this meter which upon turn on maintains a very small biasing current to the electrode for a period of seven days This keeps the electrode polarized and enables the user to make immediate measurements within a 7 day period without waiting for the electrode to re polarize Each time the ExStik is turned on the polarization timer i...

Страница 4: ...s 8 Optional zero oxygen calibration improves measurement accuracy for very low or very high DO measurements Place the electrode in a zero oxygen calibration solution such as 5 sodium sulfite wait for stability and press the CAL RECALL button until CAL is shown in the in the lower display Stability in a zero solution may take many minutes depending on electrode history Note Sodium Sulfite can beco...

Страница 5: ...ure between the electrode and the solution the longer it will take for the reading to stabilize Stabilization time can vary from thirty 30 seconds to five 5 minutes Measurement Units The meter can be set to measure either saturation mg l or parts per million ppm To change the mode 1 Press and hold the MODE HOLD button for 2 seconds and the display will begin to scroll through the different units o...

Страница 6: ...off and on again using the ON OFF button Low Battery Indication When the battery voltage falls below the operating threshold BAT will appear on the display Refer to the Maintenance section for battery replacement information Storing Readings 1 Press the MODE HOLD button to store a reading The storage location number will be displayed on the lower display while the main display shows the stored rea...

Страница 7: ...te If you are measuring DO in a small container the probe will consume oxygen as it is measuring and the measurement value will continue to go lower If the unit appears to be locked display frozen It is possible that the Data Hold mode has been inadvertently accessed by a press of the MODE HOLD button HOLD will be displayed in the bottom left of the LCD Simply press the MODE HOLD button again or t...

Страница 8: ...is obligated to take end of life devices to a designated collection point for the disposal of electrical and electronic equipment Other Battery Safety Reminders o Never dispose of batteries in a fire Batteries may explode or leak o Never mix battery types Always install new batteries of the same type Electrode Replacement 1 To remove the electrode first turn the instrument OFF and then unscrew and...

Страница 9: ...ive gold cathode 6 Set the new replacement bonded cap on a flat surface Leave the cap in this position throughout the replacement process 7 Fill the bonded cap with the electrolyte solution up to the bottom of the threads on the inside of the cap 8 Tap the side of the bonded cap to help jar free any trapped air bubbles from the electrolyte solution 9 Keeping the cap in a fixed position on a flat s...

Страница 10: ...in 300m 1000 ft increments Measurement Storage 25 tagged numbered data sets with recall Battery Power Four 4 CR2032 button batteries Low Battery Indication BAT appears on the LCD Auto Power Off After 10 minutes of inactivity APO override available Dimensions Weight 36 x 173 x 41mm 1 4 x 6 8 x 1 6 110g 3 8 oz Measurement Range Resolution Accuracy Saturation 0 to 200 0 0 1 2 0 FS full scale Dissolve...

Страница 11: ...ng surface requiring constant sample movement across membrane Move probe in sample or stir sample Slow Response Dirty or damaged membrane Replace electrolyte and membrane cap Electrode cannot be calibrated Depleted electrolyte Dirty or damaged membrane Replace electrolyte and membrane cap Electrode cannot be calibrated after replacing the electrolyte and membrane cap Dirty probe Cathode is not a s...

Страница 12: ...y in succession Displays SAL Changing Salinity Compensation On SAL Momentary press of the MODE HOLD button each button press increases the ratio by 1 ppt part per thousand the value cycles from 0 to 50 Exit Salinity Compensation On SAL Press and release CAL RECALL button for 2 seconds to enter Altitude Compensation Or Press and release Cal Button one more time to enter measurement mode It is neces...

Страница 13: ...ne Caps 15 mL KCL Filling Solution Polishing Paper EX010 Extension Cable 3 feet 1 Meter Probe Weight EX050 Extension Cable 16 feet 5 Meter Probe Weight DO610 ExStik II DO pH Conductivity Meter Kit Contains EC500 pH Conductivity Salinity TDS ExStik II Meter DO600 Dissolved Oxygen ExStik II Meter Single use pH buffer pouches 4 7 and 10pH Sample Cups with Cap Weighted Base for Sample Cups and Batteri...

Отзывы: