DO600-en-EU_V7.2 09/14
11
Appendices
Troubleshooting
Guide
Symptom
Possible
Causes
Action
Unit
will
not
power
on
‐
Batteries
not
in
place
‐
Dead
batteries
‐
Battery
polarity
incorrect
‐
Replace
Batteries
‐
Replace
Batteries
‐
Reorient/Replace
Batteries
“BAT”
indicator
shown
on
display
‐
Batteries
are
weak
‐
Replace
Batteries
Unstable
Readings
‐
Insufficient
electrolyte
in
probe
(Air
bubbles
present
when
probe
is
inverted.)
‐
Electrolyte
is
depleted
‐
Replace
electrolyte,
and
membrane
cap
assembly.
Readings
drift
down
‐
Insufficient
stirring
(the
probe
consumes
oxygen
at
the
measuring
surface,
requiring
constant
sample
movement
across
membrane)
‐
Move
probe
in
sample
or
stir
sample
Slow
Response
‐
Dirty
or
damaged
membrane
‐
Replace
electrolyte
and
membrane
cap
Electrode
cannot
be
calibrated
‐
Depleted
electrolyte
‐
Dirty
or
damaged
membrane
‐
Replace
electrolyte
and
membrane
cap
Electrode
cannot
be
calibrated
after
replacing
the
electrolyte
and
membrane
cap
‐
Dirty
probe
(Cathode
is
not
a
shiny
gold
color)
‐
Clean
cathode
with
Polishing
paper
*
perform
the
electroninc
Zero
calibration
Sample
reading
is
frozen
‐
Unit
is
in
“HOLD”
mode
‐
Unit
is
locked
‐
Release
HOLD
(momentarilly
press
Mode/Hold
button)
‐
Remove
batteries,
press
ON/OFF
button,
replace
batteries
and
restart
* Polishing
paper
is
available
in
the
DO603
membrane
kit
netzerotools.com
netzerotools.com