47
6 Using the Echo-to-Echo Option (TKG150
only)
The
Echo-to-Echo
option allows you to make
measurements between two consecutive back-wall
echoes. Therefore, a good usage of the
Echo-to-Echo
option is for measuring through coatings to measure only
the true metal thickness. There are two Echo-to-Echo
options you can select, which are as follows:
Echo-to-Echo On: Enables the Echo-to-Echo
function
Echo-to-Echo Off: Disables the Echo-to-Echo
function
In the
Measurements
mode screen in the Main menu,
highlight the
E-TO-E
option and select the desired
option using the right/left arrow keys. Press F1 to exit
Measurement mode and save settings and F1 again to get
back to the measurement mode.