background image

Model TK30  Version 1.2  July 2002 

2

Contents of Kit 

  Model MT330 True RMS MultiMeter 

  Model CA330 300A AC Current Clamp 

  Model 480172 AC Line Splitter 

  Model 40130 Voltage Detector 

  Set of 1000V Category III test leads 

  Type K thermocouple temperature probe 

  Model CA900 Carrying Case 

  Operation Manuals for the Models MT330, CA300, and TK30 

 
 
 

Current Clamp with MultiMeter 

The Current Clamp connects to the MultiMeter as shown and can be used to measure 
current for the ranges listed in the Current Clamp’s specifications. 

 

Operation 

1.  Connect the Current Clamp test 

leads to the COM and V inputs of 
the MultiMeter. 

2.  Set the MultiMeter to the 

function 

range 

3.  Press the DC/AC button to select 

AC (“

AC

” appears in the display). 

4.  Press the RANGE button until the 

milliVolt range is selected (“

mV

” 

appears in the display). 

5.  If the anticipated current is less than 10A, zero the meter using the 

REL

 button. (see 

zeroing note) 

6.  Clamp around the conductor under test and read the current measurement on the 

meter’s LCD. 

7.  The Clamp outputs 1mV AC for each Ampere of current measured. 

Note:

 The zero offset voltage is approximately 5mV and this voltage is not linear. The 

offset will affect low current measurements, but will have little affect on high current 
measurements. For the best measurement accuracy, press the Relative button to 
zero the meter for currents <10A. Do not zero the meter for currents higher than 
10A. 

 

Interpreting the Display 

Display Current 

005.0 mV 

5 amps 

020.0 mV 

20 amps 

250.0 mV 

250 amps 

 

3.999

300
AMP

AC
CL

AMP

ADAPTOR

OUTP

UT

VOL
TA

GE:
1mV

AC

Per

1A

AC

CA
TII
600V

CA300

Отзывы: