SETTING THE RECORDING SAMPLE RATE
The Sampling Rate is the rate at which the RH520 automatically records measurements
1. Press
the
SET
,
VIEW
, and
TIME
arrow buttons simultaneously from the Standard View
2. The current Sample Rate (in minutes) will appear in the numerical temperature display
area
3. The whole minutes area of the numeric display will flash
4. The MIN and the SET icons will switch on. All other TEMP and RH indicators and
numeric displays will switch off
5. The projected TIME and DAYS representing the moment the internal memory will be
full (based on the sample rate) is displayed in the TIME and DATE display areas.
6. Use
the
UP-DOWN
arrow buttons to increment/decrement the rate. Use the
LEFT-
RIGHT
arrow buttons to step through the units of time
7. Sample rates can be set from 0.1 minutes up to 199.9 minutes
8. Press
the
SET
button at any time to save changes and return to the Standard View
9. Press
VIEW
at any time to return to the Standard View without saving the changes
SETTING THE TEMPERATURE ALARM LIMITS
1. Press
the
SET
,
TEMP
, and
ALARM
buttons simultaneously from the Standard View
2. The ALARM, SET, and MAX indicators will switch on. All of the RH indicators will
switch off
3. Use
the
UP-DOWN
arrow buttons to increment/decrement the HIGH temperature
alarm limit. Use the
LEFT-RIGHT
arrow buttons to step through the decades
4. Press
the
ALARM
button. The MIN (low alarm) indicator will switch on
5. Use
the
UP-DOWN
arrow buttons to increment/decrement the LOW temperature alarm
limit. Use the
LEFT-RIGHT
arrow buttons to step through the decades
6. The HIGH and LOW alarm limits cannot overlap
7. Use
the
ALARM
button to toggle between the HIGH and the LOW alarm limits
8. Press
the
SET
button at any time to save changes and return to the Standard View
9. Press
VIEW
at any time to return to the Standard View without saving the changes
SETTING THE RH ALARM LIMITS
1. Press
the
SET
,
RH
, and
ALARM
buttons simultaneously from the Standard View
2. The ALARM, SET, and MAX indicators will switch on. All of the TEMP indicators will
switch off
3. Use
the
UP-DOWN
arrow buttons to increment/decrement the HIGH RH alarm limit.
Use the
LEFT-RIGHT
arrow buttons to step through the decades
4. Press
the
ALARM
button. The MIN (low alarm) indicator will switch on
5. Use
the
UP-DOWN
arrow buttons to increment/decrement the LOW RH alarm limit.
Use the
LEFT-RIGHT
arrow buttons to step through the decades
6. The HIGH and LOW alarm limits cannot overlap
7. Use
the
ALARM
button to toggle between the HIGH and the LOW alarm limits
8. Press
the
SET
button at any time to save changes and return to the Standard View
9. Press
VIEW
at any time to return to the Standard View without saving the changes