RH350/355    V1.7    11/04 

RS-232 PC Communications Interface 

RS-232 data format 

Baud rate; 9600, Parity; none, Data bits; 8, Stop bits; 1, Flow control; none 

RS-232 Connection 

Connect the optional interface cable from the meter to the communications port of the PC. 

Using Windows® Terminal or HyperTerminal program for PC Communication 

The Terminal or the HyperTerminal program is included in all Windows® operating systems. 
Typically it is located using the following path: 

START Menu > Programs > Accessories > Communications > Terminal or HyperTerminal 
When in the Terminal or HyperTerminal program, configure as follows: 

1. In 

the 

Connection Description

 window, enter a name, select an icon and click OK.  

2. In 

the 

Connect To

 window, select COM1 or COM2 (depending on which port is being 

used) and click OK. 

3. In the 

Port Settings

 window, select the following data format: Baud rate 9600, no 

parity, 8 data bits, 1 stop bit, and no flow control. Click OK to enter the main 
HyperTerminal window. 

4. In the 

File Menu

, select 

Properties > Settings > ASCII SETUP

, select “echo typed 

characters” and “Send line ends with line feeds”. Click OK twice to return to the main 
HyperTerminal window. 

Format 

Txxx.xC:Hxx.x%:Txxx.xC:Txxx.xC:Txxxx.xC:Txxxx.xCLRCCRLF or 
Txxx.xF:Hxx.x%:Txxx.xF:Txxx.xF:Txxxx.xF:Txxxx.xFLRCCRLF 
 
Where: The 1

st

 value is Air Temperature, the 2

nd

 value is Humidity, the 3

rd

 value is Dew 

Point, the 4

th

 value is Wet Bulb, the 5

th

 value is T1, the 6

th

 value is T2. 

Format for error value: Exxn xx is error code, the unit is n 

For example: if T1 is disconnected, then transmit: 
Txxx.xC:Hxx.x%:Txxx.xC:Txxx.xC:TE04n:Txxxx.xCLRCCRLF or 
Txxx.xF:Hxx.x%:Txxx.xF:Txxx.xF:TE04n:Txxxx.xFLRCCRLF 

Error Codes 

An error message will appear on the display if the meter fails an internal diagnostic test. 

Note

: Error messages E4 and E7 will not display on the LCD. They will be transmitted to the 

PC via the RS-232 cable and viewed in HyperTerminal. 

1. 

E2

: Air temperature is over the range. 

2. 

E3

: Air temperature is below the range.  

3. 

E4

: T1 Thermocouple not plugged in 

4. 

E5

: T1 temperature is over the range. 

5. 

E6

: T1 temperature is below the range. 

6. 

E7

: T2 thermocouple not plugged in 

7. 

E8

: T2 temperature is over the range. 

8. 

E9

: T2 temperature is below the range. 

9. 

E11

: Temperature of isothermal block is over the range. 

10. 

E12

: Temperature of isothermal block is below the range. 

11. 

E21~E23

: Circuit error. Repair/replacement necessary 

12. 

E25

: Circuit error. If problem persists, repair/replacement necessary 

13. 

E26

: Circuit error. If problem persists, repair/replacement necessary 

Содержание RH355

Страница 1: ...perature T1 T2 Air T1 T1 DP Humidity Air Temperature Dew Point and Wet Bulb Advanced features include built in memory and recall of up to 99 data points Min MAX Average readings from the captured data...

Страница 2: ...lculated from RH and Air temperature measurements Low Battery Indicator Yes Overrange Indicator E2 air E5 T1 E8 T2 Underrange Indicator E3 air E6 T1 E9 T2 Operating Conditions 4 to 122 o F 20 to 50 o...

Страница 3: ...ween Air temperature T1 temperature Air T1 and DP Dew Point 7 Pressing the T2 WB button toggles the lower left display between WB Wet Bulb T1 T2 T2 and T1 DP Dew Point Selecting temperature units of m...

Страница 4: ...enter auto power off settings Press the MEM button to select non sleep n or 2 5 10 20 40 and 60 minutes autosleep Press T2 WB to confirm the selection Manual Datalogging Manual datalogging allows the...

Страница 5: ...reading may vary depending on operating temperature will flash on the LCD screen After 30 minutes the flashing will stop to indicate that the 33 calibration is complete Note To exit from the calibrati...

Страница 6: ...x xC Hxx x Txxx xC Txxx xC Txxxx xC Txxxx xCLRCCRLF or Txxx xF Hxx x Txxx xF Txxx xF Txxxx xF Txxxx xFLRCCRLF Where The 1 st value is Air Temperature the 2 nd value is Humidity the 3 rd value is Dew P...

Страница 7: ...ee of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies on sensors and cables If it should become necessary to return the instrument for service...

Отзывы: