10
PRC20-EU-EN-V1.2 1/13
Specifications
General Specifications
Display
Dot matrix LCD
Cold junction compensation
0.02°F per °F (0.03°C per °C)
Thermocouple Standard and Scale
NIST 175, ITS-90
Meter Power
6 AA batteries or AC adaptor
Auto Power OFF
Meter automatically powers off after 10 minutes of inactivity
Operating Temperature
5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature
-20ºC to 60ºC (- 4ºF to 140ºF)
Operating Humidity
Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at
104ºF (40ºC)
Storage Humidity
<80%
Operating Altitude
2000 meters (7000ft) maximum
Dimensions
159 x 80 x 44 mm (6.3 x 3.2 x 1.7")
Weight
225g (8 oz.) without batteries
Range Specifications
Source and Measure Ranges
Resolution
Accuracy
(%
of reading)
Type J
-58 to 1832
°
F (-50 to 1000
°
C)
Type K
-58 to 2498
°
F (-50 to 1370
°
C)
Type T
-184 to 752
°
F (-120 to 400
°
C)
Type E
-58 to 1382
°
F (-50 to 750
°
C)
Type C
32 to 3182
°
F (0 to 1750
°
C)
Type R
32 to 3182
°
F (0 to 1750
°
C)
Type S
32 to 3182
°
F (0 to 1750
°
C)
Type N
-58 to 2372
°
F (-50 to 1300
°
C)
0.1
°
(measure)
1
°
(source)
± (0.05% + 1
°
C /1.8
°
F)
mV
Measure: -10mV to 60mV
Source: -5mV to 55mV
0.01mV
± (0.01% + 1digit)
Copyright
©
2013 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
www.extech.com
www.
GlobalTestSupply
.com
Find Quality Products Online at: