5
AC & AC+DC VOLTAGE
Accuracy*
MM570 MM560
Range
20Hz to 45Hz
500.00mV,
5.0000V, 50.000V
1.5% + 40d
500.00V, 1000.0V unspecified
unspecified
45Hz
to
300Hz
500.00mV 0.3%
+
20d
5.0000V, 50.000V 0.8% + 20d
500.00V, 1000.0V 0.4% + 40d
0.8% + 60d
300Hz
to
5kHz
300Hz to 1kHz
500.00mV
0.3% + 10d
0.8% + 40d
5.0000V, 50.000V,
500.00V
0.4% + 40d
2.0% + 60d
1000.0V 0.8%
+
40d
(300Hz-1kHz)
1.0% + 40d
5kHz
to
20kHz
1kHz to 20kHz
500.00mV 0.5%
+
20d
1dB**
5.0000V, 50.000V 0.8% + 20d
2dB**
500.00V 0.5%
+
20d
3dB**
1000.0V Unspecified
Unspecified
20kHz
to
100kHz
500.00mV 2.0%
+
40d
5.0000V, 50.000V 4.0% + 40d**
500.00V, 1000.0V Unspecified
Unspecified
*From 5% to 10% of range: accuracy % of reading (or in dB) + 80digits
**From 5% to 10% of range: accuracy % of reading (or in dB) +180digits
**From 10% to 15% of range: accuracy % of reading (or in dB) +100digits
CMRR: >90dB @ DC to 60Hz, Rs=1k
Ω
Input Impedance: 10M
Ω
, 30pF nominal (80pF nominal for 500mV range)
Residual reading less than 50 digits with test leads shorted