background image

MN25 Version 2.2 Aug 2003 

9

AC Voltage Measurements 

WARNING: 

 Risk of Electrocution. The probe tips may not be 

long enough to contact the live parts inside some 240V outlets 
for appliances because the contacts are recessed deep in the 
outlets. As a result, the reading may show 0 volts when the 
outlet actually has voltage on it. Make sure the probe tips are 
touching the metal contacts inside the outlet before assuming 
that no voltage is present. 
 

CAUTION: 

 Do not measure AC voltages if a motor on the 

circuit is being switched ON or OFF. Large voltage surges may 
occur that can damage the meter. 

1.  Set the function switch to the highest V

AC

 

position. 

2.  Insert the black test lead banana plug into 

the negative (COM) jack. 
Insert red test lead banana plug into the 
positive (V) jack. 

3.  Touch the black test probe tip to the 

negative side of the circuit.  
Touch the red test probe tip to the positive 
side of the circuit. 

4.  Read the voltage in the display. Reset the function switch to 

successively lower V

AC

 positions to obtain a higher 

resolution reading. The display will indicate the proper 
decimal point and value. 

 

Содержание MN25

Страница 1: ...User s Guide MiniTecTM Series Model MN25 MultiMeter...

Страница 2: ...r packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenanc...

Страница 3: ...0 seconds max every 15 minutes Resistance Continuity 250Vrms for 15sec max 2 USE EXTREME CAUTION when working with high voltages 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 5...

Страница 4: ...N symbol indicates a potentially hazardous situation which if not avoided may result damage to the product This symbol advises the user that the terminal s so marked must not be connected to a circuit...

Страница 5: ...0 reading 2 digits 200mA 100 A 1 8 reading 5 digits AC Current A AC 10A 10mA 3 0 reading 5 digits 200 0 1 2000 1 20k 0 01k 200k 0 1k 0 8 reading 2 digits Resistance 2000k 1k 1 0 reading 2 digits 9V 10...

Страница 6: ...al Low Battery Indication BAT is displayed if battery voltage drops below operating voltage Battery 9 Volt Battery Fuses mA A ranges 0 2A 250V fast blow A range 10A 250V fast blow Operating Temperatur...

Страница 7: ...e jack Note Tilt stand fuse and battery compartment are on rear of unit LCD Display Indicators Continuity Diode test m milli volts amps k kilo ohms ohms V DC volts direct current V AC volts alternatin...

Страница 8: ...roper measurement when connected to a circuit DC Voltage Measurements CAUTION Do not measure DC voltages if a motor on the circuit is being switched ON or OFF Large voltage surges may occur that can d...

Страница 9: ...N Do not measure AC voltages if a motor on the circuit is being switched ON or OFF Large voltage surges may occur that can damage the meter 1 Set the function switch to the highest VAC position 2 Inse...

Страница 10: ...and insert the red test lead banana plug into the mA jack 3 For current measurements up to 10A DC set the function switch to the 10A range and insert the red test lead banana plug into the 10A jack 4...

Страница 11: ...ition and insert the red test lead banana plug into the mA jack 3 For current measurements up to 10A AC set the function switch to the 10A range and insert the red test lead banana plug into the 10A j...

Страница 12: ...of the circuit will not interfere with the resistance reading 4 Read the resistance in the display and then set the function switch to the lowest position that is greater than the actual or any antic...

Страница 13: ...orted devices will indicate near 0mV and an open device will indicate I in both polarities Continuity Check WARNING To avoid electric shock never measure continuity on circuits or wires that have volt...

Страница 14: ...ry into battery holder observing polarity 5 Put the battery door back in place Secure with the screw Fuse Replacement WARNING To avoid electric shock disconnect the test leads from any source of volta...

Страница 15: ...bration services for the products we sell Extech also provides NIST certification for most products Call the Customer Service Department for information on calibration services available for this prod...

Отзывы: