LCR200-EU-EN V1.5 11/13
Range
Accuracy
Accuracy
10kHz
100kHz
20
Ω
± ( 1.0%rdg + 5digits )
± ( 2.0%rdg + 5digits )
After calibration
200
Ω
± ( 0.5%rdg + 5digits)
± ( 0.5%rdg + 5digits )
2k
Ω
± ( 0.5%rdg + 5digits )
± ( 0.5%rdg + 5digits )
20k
Ω
± ( 0.5%rdg + 5digits )
± ( 0.5%rdg + 5digits )
200k
Ω
± ( 0.5%rdg + 5digits )
± ( 0.5%rdg + 5digits )
2M
Ω
± ( 1%rdg + 5digits )
± ( 2.0%rdg + 5digits )
After calibration
20M
Ω
± ( 2.0%rdg + 5digits )
---------------------
After calibration
* If the impedance is larger than 10k
Ω
, Rp is shown on the display.
* If the impedance is less than 10k
Ω
, Rs is shown on the display.
Capacitance ( Cp/Cs ) : D
≦
0.1
Range Accuracy
Accuracy
Remark
100Hz/120Hz
1000Hz
20pF
± (2.0%rdg + 5digits )
± (1.0%rdg + 5digits )
After calibration
200pF
± (1.0%rdg + 5digits )
± (1.0%rdg + 5digits )
After calibration
2000pF
± (0.8%rdg + 5digits )
± (0.8%rdg + 5digits )
After calibration
20nF
± (0.5%rdg + 5digits )
± (0.5%rdg + 5digits )
200nF
± (0.5%rdg + 5digits )
± (0.5%rdg + 5digits )
2000nF
± (0.5%rdg + 5digits )
± (0.5%rdg + 5digits )
20uF
± (0.5%rdg + 5digits )
± (0.5%rdg + 5digits )
200uF
± (0.5%rdg + 5digits )
± (0.5%rdg + 5digits )
After calibration
2000uF
± (1.0%rdg + 5digits )
± (1.0%rdg + 5digits )
After calibration
20mF
± (2.0%rdg + 5digits )
------------------------
After calibration
Capacitance ( Cp/Cs ) : D
≦
0.1
Range Accuracy
Accuracy
Remark
10kHz
100kHz
20pF
± (1.0%rdg + 5digits )
± (1.0%rdg + 5digits )
After calibration
200pF
± (0.5%rdg + 5digits )
± (0.5%rdg+ 5digits )
After calibration
2000pF
± (0.5%rdg + 5digits )
± (0.5%rdg+ 5digits)
After calibration
20nF
± (0.5%rdg + 5digits )
± (0.5%rdg+ 5digits )
200nF
± (0.5%rdg + 5digits )
± (0.5%rdg+ 5d igits)
2000nF
± (0.5%rdg + 5digits )
± (0.5%rdg+ 5digits )
20uF
± (0.8%rdg + 5digits )
± (0.8%rdg+ 5digits )
200uF
± (1.0%rdg + 5digits )
--------------------------
After calibration
* If the impedance is larger than 10k
Ω
, Cp is shown on the display.
* If the impedance is less than 10k
Ω
, Cs is shown on the display.