EC400-en-GB_V2.7 12/16
11
Cleaning Recommendations
When cleaning the probe, take care not to scratch or damage the platinized electrode surfaces.
Contaminant
Cleaning Solution
Instructions
Water soluble substances
Deionized water
Soak or scrub gently with a
soft brush. Rinse thoroughly
with DI water, and dry
Grease & Oil
Warm water and household
detergent
Soak or scrub with a soft
brush, maximum of 10
minutes. Rinse thoroughly
with DI water, and dry
Heavy grease & Oil
Alcohol
Maximum of 5 minute soak,
scrub with a soft brush.
Rinse thoroughly with DI
water, and dry
Lime and hydroxide
coatings
10% acetic acid
Soak until coating dissolved,
maximum of 5 minutes. .
Rinse thoroughly with DI
water, and dry.
Troubleshooting
Problem
Possible Cause
Action
Reading is frozen
Unit is in “HOLD” mode
Press MODE/HOLD key to
exit “HOLD” mode
“BAT” message
Batteries are low
Replace batteries
Unit will not calibrate in
conductivity mode
Contaminated conductivity
standards
Use fresh standards
Unit will not calibrate in
conductivity mode
Dirty probe
Clean conductivity probe
(See cleaning instructions)
Unit will not calibrate in
conductivity mode
Damaged conductivity probe
Replace electrode
Unit will not calibrate in
conductivity mode
Trapped air bubbles
Tap or stir to release air
bubbles
Unit will not turn on
Batteries are low or dead
Replace batteries
Unit will not turn on
Batteries installed with
incorrect polarity
Replace batteries,
observe polarity
Unit will not respond to
any key presses
Internal fault
Perform hard reboot:
Remove batteries, hold
ON/OFF switch down for 5
seconds, replace batteries