CAPACITANCE MEASUREMENTS 

WARNING:

 To avoid electric shock, disconnect power 

to the unit under test and discharge all capacitors 
before taking any capacitance measurements. 
Remove the batteries and unplug the line cords. 

1. 

Set the rotary function switch to the CAP 
position. 

2. 

Insert the black test lead banana plug into the 
negative 

COM

 jack. 

3. 

Insert the red test lead banana plug into the 
positive 

CAP

 jack. 

4. 

Touch the test leads to the capacitor to be 
tested.  

5. 

The test may take up to 3 minutes or more for 
large capacitors to charge. Wait until the 
readings settle before ending the test. 

6. 

Read the capacitance value in the display 

 
 
 

FREQUENCY MEASUREMENTS 

1. 

Use the MODE button to view the 

Hz

 unit of 

measure on the LCD display. 

2. 

Insert the black test lead banana plug into the 
negative 

COM

 jack and the red test lead 

banana plug into the positive 

Hz 

jack. 

3. 

Touch the test probe tips to the circuit under 
test. 

4. 

Read the frequency on the display. 

% DUTY CYCLE 

1.  Set the rotary function switch to the 

Hz/%

 

position. 

2.  Use the MODE button to view the 

%

 unit of 

measure on the LCD display. 

3.  Insert the black lead banana plug into the 

negative 

COM

 jack and the red test lead banana 

plug into the positive 

Hz

 jack. 

4. 

Touch the test probe tips to the circuit under test. 

5. 

Read the % duty cycle on the display. 

 

 

CONTACT TEMPERATURE MEASUREMENTS 

1. 

Set the function switch to the ºF or ºC position. 

2. 

Insert the Temperature Probe into the input jacks, making 
sure to observe polarity. 

3. 

The probe can be pressed against a device under test to 
read its temperature or the probe can be held in air to read 
the ambient temperature. Allow 30 seconds for the display 
to stabilize. 

4. 

Read the temperature in the display. 

 

Содержание EX330

Страница 1: ...User s Guide Mini Multimeter with Non Contact Voltage Detector NCV Model EX330...

Страница 2: ...nals are energized This symbol indicates that a device is protected throughout by double insulation or reinforced insulation SAFETY INSTRUCTIONS This meter has been designed for safe use but must be o...

Страница 3: ...equency and temperature functions 11 RANGE button 12 HOLD button 13 Protective rubber holster must be removed to access the rear battery compartment Display Symbols and Annunciators Continuity Diode t...

Страница 4: ...of AC voltage from 100 to 600VAC simply by being held very near to a voltage source WARNING Test the AC voltage detector on a known live circuit before each use WARNING Before using the meter in the A...

Страница 5: ...the negative COM jack Insert red test lead banana plug into the positive V jack 3 Touch the black test probe tip to the neutral side of the circuit 4 Touch the red test probe tip to the hot side of c...

Страница 6: ...to 400mA set the function switch to the mA position and insert the red test lead banana plug into the mA A jack 4 For current measurements up to 10A set the function switch to the 10A range and inser...

Страница 7: ...sconnect one side of the circuit under test so the rest of the circuit will not interfere with the resistance reading 4 Read the resistance in the display CONTINUITY CHECK WARNING To avoid electric sh...

Страница 8: ...jack and the red test lead banana plug into the positive jack 3 Use the MODE button to view the icon on the display 4 Touch the test probes to the diode under test Forward voltage will typically indi...

Страница 9: ...2 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz jack 3 Touch the test probe tips to the circuit under test 4 Read the frequen...

Страница 10: ...EL indicator will appear on the display 3 The display will now indicate the difference between stored value and subsequent measurements 4 Press the REL button to exit the relative mode Note The Relati...

Страница 11: ...mpartment cover 7 Place the protective rubber holster on the meter WARNING To avoid electric shock do not operate the meter until the batteries and the fuses are in place and fastened securely 1 Remov...

Страница 12: ...has been evaluated for the accuracy of its readings AUTO POWER OFF The meter will automatically turn off after 15 minutes of inactivity This will conserve battery energy To turn the meter on after an...

Страница 13: ...gits 4V 0 001mV 40V 0 01V 400V 0 1V 1 5 reading 3 digits AC Voltage V AC 50 60Hz 600V 1V 2 0 reading 4 digits 400 A 0 1 A 1 0 reading 3 digits 4000 A 1 A 40mA 0 01mA 400mA 0 1mA 1 5 reading 3 digits D...

Страница 14: ...e and 100kHz 0 1 to 99 9 0 1 1 2 reading 2 digits Duty Cycle For duty cycle the pulse width range is 100 s 100ms Frequency 5Hz to 150kHz 4 to 1382 F 1 F Temp type K 20 to 750 C 1 C 3 0 reading 8 digit...

Страница 15: ...or indoor use and protected against the users by double insulation per EN61010 1 and IEC61010 1 2nd Edition 2001 to CAT II 1000V CAT III 600V Pollution Degree 2 The meter also meets UL 61010 1 Second...

Страница 16: ...e sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misus...

Отзывы: