Operation
Measurements
Slide
the
power
switch
to
the
ON
position.
The
sensor
is
located
on
the
rear
of
the
meter
2
inches
(5
cm)
from
the
top
of
the
meter
on
the
plane
indicated
by
the
left
and
right
arrows
next
to
the
word
SENSOR
on
the
front
of
the
meter.
Touch
the
TOP
of
the
meter
to
the
microwave
door
and
various
spots
on
the
microwave
enclosure.
The
display
will
indicate
the
EMF
value
that
the
sensor
is
reading.
If
the
meter
senses
a
reading
outside
the
specified
range
of
the
instrument
the
‘OL’
(overload)
display
symbol
will
appear
on
the
LCD.
Alarm
Limit
When
the
measurement
is
>1mW/cm^2
the
meter’s
buzzer
will
sound
to
alert
the
user
that
the
EMF
value
exceeds
the
alarm
limit.
Peak
Hold
To
capture
the
highest
value
on
the
LCD
display,
momentarily
press
the
Peak
Hold
key.
The
PEAK
display
icon
will
appear
on
the
display
and
the
meter
will
only
display
the
highest
reading.
Momentarily
press
the
Peak
Hold
key
again
to
return
to
normal
operation
(the
PEAK
display
icon
will
switch
OFF).
Microwave
Oven
Safety
Standards
In
the
U.S.,
please
check
with
The
Food
and
Drug
Administration
(FDA)
for
microwave
oven
safety
standards
with
regard
to
leakage
requirements.
The
microwave
leakage
limit
in
the
U.S.
at
the
time
of
this
writing
is
5mW/cm
2
at
2
inches
(5cm).
This
may
differ
from
country
to
country
or
may
have
been
updated
in
the
U.S.
since
the
time
of
this
writing.
Be
sure
to
obtain
the
appropriate
and
current
standards
and
regulations
for
the
appropriate
country
or
location
before
using
this
device.
Neither
FLIR
Systems
nor
Extech
Instruments
is
responsible
for
damage
to
equipment
or
injury
to
persons
with
regard
to
the
use
of
this
meter
or
in
the
interpretation
of
the
safety
standards
in
question.
Full
safety
responsibility
is
assumed
by
the
user
at
all
times.
Battery
Replacement
When
the
battery
icon
appears
on
the
LCD
or
if
the
meter
does
not
switch
ON,
the
9V
battery
must
be
replaced.
The
battery
compartment
is
located
on
the
rear
of
the
instrument.
Open
the
compartment
to
replace
or
install
the
battery
and
then
secure
the
compartment
before
switching
the
meter
ON.