Trouble Shooting Guide
Problem Cause Remedy
Unit does not turn
on
Batteries are flat Recharge batteries or
run on mains
shown on
display, or unit
unexpectedly
turns off
Low batteries
Recharge batteries
O
2
reading “
Lo
”
after stabilising in
clean air
O
2
sensor failed
Contact Extech
“
High
” shown on
display
Over-range
reading
Remove probe from
gas source
immediately and allow
unit to run in clean air
for 5 minutes
Ratio reads “
rnG
”
CO or CO2
reading is zero.
Unable to
calculate
Probe is not in the
flue/appliance not
running
Battery life shorter
than expected
Back light or
leak detector
has been used
excessively
Turn off back light
when not required
“
CAL
” shown on
display
Calibration
required
Contact Extech
“
Prob
” displayed
Instrument
calibrated whilst
probe in flue
Turn off instrument.
Turn back on with
probe outside flue
Continuous alarm
sounds
Excessive levels
of CO
Remove probe from
flue, and allow unit to
run in clean air for 5
minutes