background image

User Manual 

 

 

 

Coating Thickness Tester 

 

 

Model CG104 

 

 
 
 
 
 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

 

Additional User Manual Translations available at www.extech.com 

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание CG104

Страница 1: ...User Manual Coating Thickness Tester Model CG104 Additional User Manual Translations available at www extech com www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ...heck for the latest version of this User Guide Product Updates Product Registration and Customer Support Features LED backlighting Reversible LCD orientation Auto Power OFF Low Battery Indication Zero Function and Two Point Calibration Datalogging function High Low Alarm Alerts Selectable Measurement Units Supplied Zero and Standard Coating Substrates Description Meter Description 1 Ferrous non fe...

Страница 3: ... Measurement mode Backlight and ZERO buttons are disabled in the Options menu Display Icon Description Fe nFe Ferrous and Non ferrous substrates AT Auto metal sensing mode High Low Alarm indicators MEM MIN MAX AVG Recording is active CAL Calibration is active µm Micrometer unit of measure mils mils unit of measure millimeters 2 54 100 MAX MIN Maximum minus Minimum reading MIN MAX AVG Minimum Maxim...

Страница 4: ...coating Substrate zero calibration must be performed before each use The enclosed zeroing plates are suitable for the zeroing calibration as described in this User Guide To ensure measurement accuracy the Two point Calibration as described in this User Guide is recommended before each use Zeroing on a specific material substrate still needs to be done before taking formal measurements such as for ...

Страница 5: ...2 Make sure that the tip of the meter is clean of all oil and particles 3 Press the spring loaded sensor against the surface of a blank round ferrous or non ferrous substrate 4 While maintaining contact with the substrate momentarily press the ZERO CLEAR button The meter will beep once and 0 will appear on the display indicating that the meter has been zeroed 5 Place a reference film on either the...

Страница 6: ... flashing in the upper right corner of the display 3 Press MENU again and the LCD will display Lo in the upper right hand corner Use the up and down arrow buttons to set the value to 0 for an uncoated substrate or to the thickness value of the thinner reference film being used for the calibration 4 Press MENU ENTER to confirm the value The display will flash PUSH 5 Press the tip of the gauge to th...

Страница 7: ...e basics use the arrow keys to select one of the record options Max Min Max Min Avg n Saved values Press and hold the ESC button for 5 seconds 2 Press the or button to select delete NO or YES 3 To retain your saved readings Select NO and press MENU ENTER to revert to the previous setting 4 To delete all Records Select YES and press MENU ENTER The meter clear the saved values and return to the norm...

Страница 8: ...es and store them separately Battery Replacement Installation 1 Remove the Phillips head screw that secures the rear battery door 2 Open the battery compartment 3 Replace install the two 1 5V AAA batteries observing proper polarity 4 Secure the battery compartment Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batte...

Страница 9: ...ils 10mils 80mils 0 1mils 0 100μm 1μm 100μm 1020μm 1μm 0 10mils 0 1mils 10mils 40 0mils 0 1mils Basic critical thickness 0 to 80mils 0 to 2000um 0 to 40mils 0 to 1020um Industrial standards Conforms to GB T 4956 1985 GB T 4957 1985 JB T 8393 1996 JJG 889 95 and JJG 818 93 Temperature Coefficient 0 1 times the applicable accuracy specification per C from 18 C to 28 C 64 F to 82 F 1 Accuracy note Ac...

Страница 10: ...30 seconds Standby Current Consumption 6μA Battery Power 1 5V AAA x 2pcs Battery Life 17 hours continuous use with a typical alkaline battery Low Battery Indication Icon is displayed when the battery voltage drops below the operating level Dimensions 105 H x 55 W x 27 D mm 4 1 x 2 2 x 1 1 Weight Approx 80g 2 82 oz including battery Copyright 2015 2017 FLIR Systems Inc All rights reserved including...

Отзывы: