19
407790A-EN_V2.4 1/20
3.5 Selecting the Lmax/Lmin type
Band - (Band maximum/band minimum):
The analysis result applies to the point where the level for each frequency band was maximum
or minimum within the sampling period.
AP - (All-pass maximum/all-pass minimum):
The analysis result applies to the sub channel all-pass and band-pass level values taken at the
point where the all-pass level in the main channel was maximum/minimum within the
sampling period.
AP (S) - (All-pass maximum/all-pass minimum):
The analysis result applies to the band-pass level values (excluding the sub channel all-pass
levels) taken at the point where the all-pass level in the main channel was
maximum/minimum within the sampling period. The sub
channel all-pass level is taken at
the maximum/minimum point within the sampling period, independently of the main
channel level.
1. Press the MENU button to enter the menu list screen.
2. Use the
+
/
-
buttons to select “I). Lmax/Lmin Type” then press
button.
3. Use the
+
/
-
buttons to select the desired Band, AP or AP (S) type then press
button.
4. Press MENU button to return to the measurement screen.
This setting is effective in SLM mode and Octave band analyzer mode.
Содержание 407790A
Страница 7: ...7 407790A EN_V2 4 1 20 ...
Страница 14: ...14 407790A EN_V2 4 1 20 ...
Страница 16: ...16 407790A EN_V2 4 1 20 Time ...
Страница 21: ...21 407790A EN_V2 4 1 20 AP AP AP AP ...
Страница 31: ...31 407790A EN_V2 4 1 20 ...
Страница 36: ...36 407790A EN_V2 4 1 20 ...
Страница 38: ...38 407790A EN_V2 4 1 20 ...
Страница 42: ...42 407790A EN_V2 4 1 20 ...
Страница 43: ...43 407790A EN_V2 4 1 20 ...
Страница 46: ...46 407790A EN_V2 4 1 20 ...