407768   Version 2.0   06/06 

2.  The meter should read close to, or exactly, the calibrator's dB output level. Typical 

Calibrator output levels are 94dB and 114dB. 

3.  If the meter is within ± 0.2 dB of the calibrator’s output, no adjustment is necessary. 

4.  Adjust the calibration pot in the output/calibration compartment if necessary to bring the 

meter display in line with the calibrator output signal. 

Taking Measurements 

1.  Hold the meter in hand, place it on a desktop (using the rear tilt stand), or mount it on a 

tripod using the tripod mount on the rear of the meter. 

2.  Point the microphone toward the source of noise/sound to be measured. 

3.  Read the measurement, in dB units, on the LCD display. 

Data Hold 

1.  Press the HOLD button to freeze the displayed reading. The LCD will display the icon 

HOLD when Data Hold is engaged.  

2.  Press the HOLD button to de-activate this feature. The HOLD icon will extinguish. 

3.  Note that Data Hold is not available while the meter is in the Max/Min Record mode. 

Max Hold 

1.  Press the MAX HOLD button to activate this feature. The LCD will display PH when 

Max Hold is engaged. 

2.  The meter will now display only the highest reading. Each time a higher reading is 

encountered the display will update. 

3.  Press the MAX HOLD button again to de-activate this feature. The PH icon will 

extinguish. 

4.  Note that Max Hold is not available while the meter is in the Max/Min Record mode. 

Max/Min Data Recording 

The Max/Min feature stores the maximum reading and minimum reading while the user 
takes measurements. The Max and the Min readings can later be recalled. 

1.  Press the REC MAX/MIN button to activate this feature. The LCD will display the icon 

REC and the meter will begin monitoring the highest (Max) and lowest (Min) readings. 

2.  After measurements are made, press the REC button again to view the Max reading. 

The MAX icon will appear on the LCD along with the highest reading. 

3.  Press the REC to view the minimum (MIN) reading. The 'MIN' display icon will appear 

on the LCD along with the lowest reading. 

4.  To exit this mode, 

press and hold

 the REC button until the REC indicator 

extinguishes. 

5.  To clear a Max or Min reading, press the HOLD key while viewing either the MAX or 

MIN value. Note that exiting the Record mode also clears the MAX and MIN values. 

Analog Output 

The AC analog output transmits 0.5V AC rms full scale. The output impedance is 
600

Ω 

maximum. The mono 3.5mm mini-jack is located in the output/calibration 

compartment on the right side of the instrument. Use a phono plug (3.5mm mono) to 
connect to the output jack. 

Содержание 407768

Страница 1: ...User s Manual Digital Sound Level Meter Model 407768 ...

Страница 2: ...0 to 130dB Manual Ranges 30 to 80dB 50 to 100dB 80 to 130dB Frequency weighting A and C Programmable Applicable standards ANSI IEC Type 2 Resolution 0 1dB Maximum Minimum record Highest and lowest readings are stored for later recall Data Hold Displayed reading is held when HOLD key is pressed Max Hold Highest reading displayed only Response time Fast 125ms Slow 1s Programmable AC Analog output 0 ...

Страница 3: ...play on the lower left hand area of the LCD Note With A weighting selected the meter responds like the human ear boosting and cutting the noise amplitude over the frequency spectrum see Appendix A weighting is used for environmental measurements OSHA regulatory testing law enforcement and workplace design Select C weighting for flat response measurements no boost or cut C weighting is suitable for...

Страница 4: ...her reading is encountered the display will update 3 Press the MAX HOLD button again to de activate this feature The PH icon will extinguish 4 Note that Max Hold is not available while the meter is in the Max Min Record mode Max Min Data Recording The Max Min feature stores the maximum reading and minimum reading while the user takes measurements The Max and the Min readings can later be recalled ...

Страница 5: ... Instructions for use can be found on the included software CD Important Note on the Automatic Power OFF feature The 407768 Sound Level Meter has an AUTO POWER OFF feature that turns the meter off after approximately 10 minutes This feature must be disabled before starting a datalogging session to avoid having the meter automatically shut off in the middle of a session Disable it by pressing the R...

Страница 6: ... six month limited warranty applies on sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website at www extech com click on Contact Extech and go to Service Department to request an RA number A Return Authorization RA number must be...

Страница 7: ... 50 HP Siren 100 Jet take off 200 Riveting machine Subway 20 Pneumatic drill 50 Vacuum cleaner 10 Large store Small office Residential area night Whisper 5 Threshold of hearing Chain saw Boiler room Freight train 100 Speech 1 Large Office Residence Sound studio more infor for Extech 407768 Phone 01235 838 555 Email cs airconcern co uk Web www airconcern co uk Air Concern Ltd Building 173 Curie Ave...

Отзывы: